Open official notice

Cyanosure dermav laser system

Official notice

26 Aug 2026

Last checked

13 Nov 2026

RecallScope risk estimate:HighHealth Canada

An official Health Canada notice concerns Cyanosure dermav laser system. Cynosure Lutronic has identified a adverse trend of complaint reports in North America for spark and loud noise resulting in patient burns during vascular treatment with the DermaV laser system. There is a remote possibility that utilizing higher fluences at 1064 nm wavelength with the current cryogen parameters can result in a spark/loud noise being observed during treatment that could potentially result in patient burn. Published 2026-08-26.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82528

Official notice

2026-08-26

The hazard

Cynosure Lutronic has identified a adverse trend of complaint reports in North America for spark and loud noise resulting in patient burns during vascular treatment with the DermaV laser system. There is a remote possibility that utilizing higher fluences at 1064 nm wavelength with the current cryogen parameters can result in a spark/loud noise being observed during treatment that could potentially result in patient burn.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

1011600210

Other

RA RA-82528

Model numbers

1011600210

Recall Numbers

82528

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - Not applicable., Model or catalogue number - 1011600210

Hazard

Cynosure Lutronic has identified a adverse trend of complaint reports in North America for spark and loud noise resulting in patient burns during vascular treatment with the DermaV laser system. There is a remote possibility that utilizing higher fluences at 1064 nm wavelength with the current cryogen parameters can result in a spark/loud noise being observed during treatment that could potentially result in patient burn.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Cynosure Lutronic Technology.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82528

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 26 Aug 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.