Biograph lso 16 slice pet
Official notice
26 Nov 2021
Last checked
26 Nov 2021
An official Health Canada notice concerns Biograph lso 16 slice pet. There is potential patient fall hazards could occur when the PHS pallet/foot extension is improperly assembled by the operator. PHS pallet/foot extension patient fall hazards could occur when the extension is improperly used to position the patient head-first. The pallet/foot extension should only be used for positioning the patient's feet. To avoid this issue, customers are advised to ensure proper operator assembly of the PHS pallet/foot extension as instructed in the operator's manual and as depicted on the extension label. And to only use the PHS pallet/foot extension for its intended purpose of feet first imaging as instructed in the operator's manual. The operator manual cautions of the improper use of the positioning aids and notes to only use the intended purpose. The foot extension itself is labelled and instructions of proper use of extension are also provided. Recall start date: 2021-11-08 Published 2021-11-26.
What to do now
Contact the manufacturer if you require additional information.
Recall number
63652
Official notice
2021-11-26
The hazard
The hazard
There is potential patient fall hazards could occur when the PHS pallet/foot extension is improperly assembled by the operator. PHS pallet/foot extension patient fall hazards could occur when the extension is improperly used to position the patient head-first. The pallet/foot extension should only be used for positioning the patient's feet. To avoid this issue, customers are advised to ensure proper operator assembly of the PHS pallet/foot extension as instructed in the operator's manual and as depicted on the extension label. And to only use the PHS pallet/foot extension for its intended purpose of feet first imaging as instructed in the operator's manual. The operator manual cautions of the improper use of the positioning aids and notes to only use the intended purpose. The foot extension itself is labelled and instructions of proper use of extension are also provided. Recall start date: 2021-11-08
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Siemens, Biograph Lso 16 Slice Pet.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Artis icono ceiling, floor and biplane systems

Artis icono ceiling

Magnetom prisma, vida, skyra, verio systems

Artis pheno

Uroskop omnia

Atellica ch reaction ring cuvette segment

Advia® and atellica® chemistry urinary cerebrospinal fluid protein (ucfp) assay

Luminos lotus max, drf max, and agile max

Axiom vertix and axiom multix

Siemens biograph mmr system

Artis icono ceiling

Cios spin, cios flow, and cios alpha
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 26 Nov 2021. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.