Open official notice

Atellica® ch urine albumin (ualb)

Official notice

17 Jul 2026

Last checked

13 Nov 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Atellica® ch urine albumin (ualb). Siemens Healthineers has confirmed, through the investigation of customer complaints, the potential for the following issues associated with processing Atellica CH Urine Albumin (UAlb) tests on Atellica CH and Atellica CI analyzers.• Falsely depressed Atellica CH Urea Nitrogen (UN_c) results• Falsely depressed Atellica CH Enzymatic Hemoglobin A1c (A1c_E) results• Increased rate of reaction cuvette QC failuresThese issues may impact quality control (QC), calibrators, and patient results for the UN_c and A1c_E assays across all sample types. Published 2026-07-17.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82353

Official notice

2026-07-17

The hazard

Siemens Healthineers has confirmed, through the investigation of customer complaints, the potential for the following issues associated with processing Atellica CH Urine Albumin (UAlb) tests on Atellica CH and Atellica CI analyzers.• Falsely depressed Atellica CH Urea Nitrogen (UN_c) results• Falsely depressed Atellica CH Enzymatic Hemoglobin A1c (A1c_E) results• Increased rate of reaction cuvette QC failuresThese issues may impact quality control (QC), calibrators, and patient results for the UN_c and A1c_E assays across all sample types.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

11537225

Other

RA RA-82353

Model numbers

11537225

Recall Numbers

82353

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 11537225

Hazard

Siemens Healthineers has confirmed, through the investigation of customer complaints, the potential for the following issues associated with processing Atellica CH Urine Albumin (UAlb) tests on Atellica CH and Atellica CI analyzers.• Falsely depressed Atellica CH Urea Nitrogen (UN_c) results• Falsely depressed Atellica CH Enzymatic Hemoglobin A1c (A1c_E) results• Increased rate of reaction cuvette QC failuresThese issues may impact quality control (QC), calibrators, and patient results for the UN_c and A1c_E assays across all sample types.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Siemens Healthcare Diagnostics.

Source & status

Tier

beta

Category

Home & Consumer

Official classification

Type II

Recall details

Recall number

82353

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 17 Jul 2026. The next one won’t wait for you to search.