No image published for this recall
Access intact pth
Official notice
29 Nov 2024
Last checked
29 Nov 2024
An official Health Canada notice concerns Access intact pth. Beckman Coulter has determined that 91 Access Intact PTH reagent packs (lot 339071) were filled with an incorrect volume of reagent in well 0. Access Intact PTH patient and QC samples with a reagent pack that has less volume than expected may: - Recover up to 30% lower than the expected results for routine PTH, with a pack imprecision of up to 10%. - Recover up to 40% lower than the expected results for intraoperative PTH, with a pack imprecision of up to 15%. - Produce a zero or near-zero result. Falsely decreased Intact PTH results could lead to a delay in diagnosis, additional diagnostic testing, and/or inappropriate treatment. Recall Start Date: November 25, 2024 Published 2024-11-29.
What to do now
Contact the manufacturer if you require additional information.
Recall number
76621
Official notice
2024-11-29
The hazard
The hazard
Beckman Coulter has determined that 91 Access Intact PTH reagent packs (lot 339071) were filled with an incorrect volume of reagent in well 0. Access Intact PTH patient and QC samples with a reagent pack that has less volume than expected may: - Recover up to 30% lower than the expected results for routine PTH, with a pack imprecision of up to 10%. - Recover up to 40% lower than the expected results for intraoperative PTH, with a pack imprecision of up to 15%. - Produce a zero or near-zero result. Falsely decreased Intact PTH results could lead to a delay in diagnosis, additional diagnostic testing, and/or inappropriate treatment. Recall Start Date: November 25, 2024
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Lots
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Beckman Coulter, Access Intact Pth.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Unicel dxi 600 access immunoassay system-total t4 assay

Dxc 700 au and au5800 clinical chemistry analyzers

Cellmek sps

Unicel dxi 600 and 800 access immunoassay systems

Beckman coulter au immunochemistry system - bicarbonate (co2) assay

Dxc 500i clinical analyzer and dxi 9000 access immunoassay analyzer

Synchron® systems direct ldl cholesterol (ldld) reagent

Synchron systems phosphorus (phosm) reagent

Cellmek sps

Unicel dxh series instruments

Dxc 500i clinical analyzers

Unicel dxi 600 access immunoassay system- thyroglobulin assay
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 29 Nov 2024. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.