Access intact pth

Official notice

29 Nov 2024

Last checked

29 Nov 2024

Critical riskHealth Canada

An official Health Canada notice concerns Access intact pth. Beckman Coulter has determined that 91 Access Intact PTH reagent packs (lot 339071) were filled with an incorrect volume of reagent in well 0. Access Intact PTH patient and QC samples with a reagent pack that has less volume than expected may: - Recover up to 30% lower than the expected results for routine PTH, with a pack imprecision of up to 10%. - Recover up to 40% lower than the expected results for intraoperative PTH, with a pack imprecision of up to 15%. - Produce a zero or near-zero result. Falsely decreased Intact PTH results could lead to a delay in diagnosis, additional diagnostic testing, and/or inappropriate treatment. Recall Start Date: November 25, 2024 Published 2024-11-29.

What to do now

Contact the manufacturer if you require additional information.

Recall number

76621

Official notice

2024-11-29

The hazard

Beckman Coulter has determined that 91 Access Intact PTH reagent packs (lot 339071) were filled with an incorrect volume of reagent in well 0. Access Intact PTH patient and QC samples with a reagent pack that has less volume than expected may: - Recover up to 30% lower than the expected results for routine PTH, with a pack imprecision of up to 10%. - Recover up to 40% lower than the expected results for intraoperative PTH, with a pack imprecision of up to 15%. - Produce a zero or near-zero result. Falsely decreased Intact PTH results could lead to a delay in diagnosis, additional diagnostic testing, and/or inappropriate treatment. Recall Start Date: November 25, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Lots

339071

Recall Numbers

76621

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Access Intact PTH, Lot or serial number - 339071, Model or catalog number - A16972

Risk / hazard

Beckman Coulter has determined that 91 Access Intact PTH reagent packs (lot 339071) were filled with an incorrect volume of reagent in well 0. Access Intact PTH patient and QC samples with a reagent pack that has less volume than expected may: - Recover up to 30% lower than the expected results for routine PTH, with a pack imprecision of up to 10%. - Recover up to 40% lower than the expected results for intraoperative PTH, with a pack imprecision of up to 15%. - Produce a zero or near-zero result. Falsely decreased Intact PTH results could lead to a delay in diagnosis, additional diagnostic testing, and/or inappropriate treatment. Recall Start Date: November 25, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Beckman Coulter, Access Intact Pth.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type I

App estimate

critical

Recall details

Recall number

76621

Classification

Type I

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
No image published for this recall

Unicel dxi 600 access immunoassay system-total t4 assay

Medium riskHealth Canada19 Mar 2026
CA
No image published for this recall

Dxc 700 au and au5800 clinical chemistry analyzers

Medium riskHealth Canada5 Feb 2026
CA
No image published for this recall

Cellmek sps

Medium riskHealth Canada30 Dec 2025
CA
No image published for this recall

Unicel dxi 600 and 800 access immunoassay systems

Medium riskHealth Canada16 Dec 2025
CA
No image published for this recall

Beckman coulter au immunochemistry system - bicarbonate (co2) assay

Medium riskHealth Canada17 Oct 2025
CA
No image published for this recall

Dxc 500i clinical analyzer and dxi 9000 access immunoassay analyzer

Critical riskHealth Canada26 Sept 2025
CA
No image published for this recall

Synchron® systems direct ldl cholesterol (ldld) reagent

Medium riskHealth Canada25 Sept 2025
CA
No image published for this recall

Synchron systems phosphorus (phosm) reagent

Medium riskHealth Canada17 Sept 2025
CA
No image published for this recall

Cellmek sps

Medium riskHealth Canada4 Sept 2025
CA
No image published for this recall

Unicel dxh series instruments

Medium riskHealth Canada20 Aug 2025
CA
No image published for this recall

Dxc 500i clinical analyzers

Medium riskHealth Canada19 Aug 2025
CA
No image published for this recall

Unicel dxi 600 access immunoassay system- thyroglobulin assay

Medium riskHealth Canada23 Jul 2025

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 29 Nov 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.