An official Health Canada notice concerns Abl90 flex and flex plus. There is a potential risk with ABL90 Flex & ABL90 Flex Plus running RWIN10 1.0 and RWIN10 1.1. The issue is that the analyzer software may unexpectedly freeze during a sample measurement. The analyzer software includes a watchdog that detects a freeze and automatically restarts the software. If the analyzer freezes, a sample being processed will be lost. Recall Start Date: March 11, 2025 Published 2025-03-25.
What to do now
Contact the manufacturer if you require additional information.
Recall number
77178
Official notice
2025-03-25
The hazard
The hazard
There is a potential risk with ABL90 Flex & ABL90 Flex Plus running RWIN10 1.0 and RWIN10 1.1. The issue is that the analyzer software may unexpectedly freeze during a sample measurement. The analyzer software includes a watchdog that detects a freeze and automatically restarts the software. If the analyzer freezes, a sample being processed will be lost. Recall Start Date: March 11, 2025
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Product 2
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Radiometer Medical Aps, Abl90 Flex And Flex Plus.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Abl800 basic and abl8xx flex analyzers

Abl800 flex system

Harmonyair a-series surgical lighting system and alyon surgical lighting system

Quickvue® dipstick strep a test
Image: Health Canada recall noticeBissell brand steam shot™ omnireach™ and steam shot™ omni steam cleaner attachments recalled
Image: Health Canada recall noticeScubapro monorail weight pockets recalled
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 25 Mar 2025. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.