Harmonyair a-series surgical lighting system and alyon surgical lighting system
Official notice
21 May 2026
Last checked
21 May 2026
An official Health Canada notice concerns Harmonyair a-series surgical lighting system and alyon surgical lighting system. Steris identified through customer complaints that on Harmonyair A-Series Lighting System and Alyon Surgical Lighting System the plus/minus (+/-) buttons that control light intensity on the light heads may detach. If detachment occurs during a patient procedure, the buttons could enter the sterile field and disrupt the procedure resulting in a procedure delay. Users can continue to use their lighting system(s) prior to completion of the correction. Users are reminded to ensure no excess fluids remain on light head during or after cleaning, per the operator manual. Should detachment of the plus/minus (+/-) buttons occur prior to the correction being completed, users should immediately contact their local Steris representative. Published 2026-05-21.
What to do now
Contact the manufacturer if you require additional information.
Recall number
82097
Official notice
2026-05-21
The hazard
The hazard
Steris identified through customer complaints that on Harmonyair A-Series Lighting System and Alyon Surgical Lighting System the plus/minus (+/-) buttons that control light intensity on the light heads may detach. If detachment occurs during a patient procedure, the buttons could enter the sterile field and disrupt the procedure resulting in a procedure delay. Users can continue to use their lighting system(s) prior to completion of the correction. Users are reminded to ensure no excess fluids remain on light head during or after cleaning, per the operator manual. Should detachment of the plus/minus (+/-) buttons occur prior to the correction being completed, users should immediately contact their local Steris representative.
How to check if you're affected
How to check if you're affected
Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.
Skus
Other
Model Numbers
Recall Numbers
Checked manually this time?
The app watches your VIN and alerts you automatically next time — no searching, no checking back.
Turn on automatic VIN checksAffected products
Affected products
Product 1
Product 2
Product 3
Product 4
Product 5
Product 6
Product 7
Product 8
Product 9
Product 10
Product 11
Product 12
Product 13
Product 14
Product 15
Product 16
Product 17
Product 18
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected vehicles were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand
Manufacturer
Phone
Brand: Steris.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Harmonyair a-series surgical lighting system

Dual flat panel monitor (dfpm) accessory to the harmonyair® a-series and e-series surgical lighting systems

Sterile light handle cover

Solar™ gi

Omni-tract surgical retractor system

Urinary catheter trays containing cliniclean pericare wipes

Wasplab

Nephromax high pressure nephrostomy balloon catheter

Uhi-4 high flow insufflation system - insufflator

Atellica ch enzymatic hemoglobin a1c (a1c e) assay

Iceman classic3™

Kala red light face mask
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 21 May 2026. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.