Abl800 basic and abl8xx flex analyzers

Official notice

20 Jun 2024

Last checked

20 Jun 2024

Medium riskHealth Canada

An official Health Canada notice concerns Abl800 basic and abl8xx flex analyzers. Radiometer has become aware of an issue with the Abl800 series analyzers. An internal study has concluded that there is a remote probability of reporting biased out-of-specification pH results on blood samples. This may occur if the calibration solution's pH value decreases during the in-use period due to bacterial growth in the calibration solution bottles Cal1 and/or Cal2. Recall start date: June 4, 2024 Published 2024-06-20.

What to do now

Contact the manufacturer if you require additional information.

Recall number

75706

Official notice

2024-06-20

The hazard

Radiometer has become aware of an issue with the Abl800 series analyzers. An internal study has concluded that there is a remote probability of reporting biased out-of-specification pH results on blood samples. This may occur if the calibration solution's pH value decreases during the in-use period due to bacterial growth in the calibration solution bottles Cal1 and/or Cal2. Recall start date: June 4, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

75706

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Abl800 Basic And Abl8xx Flex Analyzers, Lot or serial number - All lots., Model or catalogue number - 393-801 393-800

Risk / hazard

Radiometer has become aware of an issue with the Abl800 series analyzers. An internal study has concluded that there is a remote probability of reporting biased out-of-specification pH results on blood samples. This may occur if the calibration solution's pH value decreases during the in-use period due to bacterial growth in the calibration solution bottles Cal1 and/or Cal2. Recall start date: June 4, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Radiometer Medical Aps, Abl800 Basic And Abl8xx Flex Analyzers.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

75706

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 20 Jun 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.