Open official notice

Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21, and 23

Official notice

3 Jul 2024

RecallScope risk estimate:HighFDA

Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21, and 23

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

34,102

Recall number

Z-2189-2024

Official notice

2024-07-03

The hazard

Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric shock. Health consequences may include electrical shock, burns, and cardiac arrhythmia.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

MS20407

Other

UDI-DI 04049098054447

Event Ids

94644

Model numbers

MS20407

Recall Numbers

Z-2189-2024

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21, and 23

Units affected

34,102

Hazard

Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric shock. Health consequences may include electrical shock, burns, and cardiac arrhythmia.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

AT · Austria

Affected area 5

AU · Australia

Affected area 6

BE · Belgium

Affected area 7

BG · Bulgaria

Affected area 8

BR · Brazil

Affected area 9

CA · Canada

Affected area 10

CH · Switzerland

Affected area 11

CL · Chile

Affected area 12

CN · China

Affected area 13

CR · Costa Rica

Affected area 14

CZ · Czech Republic

Affected area 15

DE · Germany

Affected area 16

EE · Estonia

Affected area 17

EG · Egypt

Affected area 18

ES · Spain

Affected area 19

FI · Finland

Affected area 20

FR · France

Affected area 21

GB · United Kingdom

Affected area 22

GR · Greece

Affected area 23

HK · Hong Kong

Affected area 24

HR · Croatia

Affected area 25

HU · Hungary

Affected area 26

IE · Ireland

Affected area 27

IL · Israel

Affected area 28

IN · India

Affected area 29

IT · Italy

Affected area 30

JP · Japan

Affected area 31

KR · South Korea

Affected area 32

KW · Kuwait

Affected area 33

LI · Liechtenstein

Affected area 34

LT · Lithuania

Affected area 35

LV · Latvia

Affected area 36

MT · Malta

Affected area 37

MX · Mexico

Affected area 38

MY · Malaysia

Affected area 39

NL · Netherlands

Affected area 40

NO · Norway

Affected area 41

NZ · New Zealand

Affected area 42

PE · Peru

Affected area 43

PK · Pakistan

Affected area 44

PL · Poland

Affected area 45

PT · Portugal

Affected area 46

RO · Romania

Affected area 47

RS · Serbia

Affected area 48

SA · Saudi Arabia

Affected area 49

SE · Sweden

Affected area 50

SG · Singapore

Affected area 51

SI · Slovenia

Affected area 52

SK · Slovakia

Affected area 53

TH · Thailand

Affected area 54

TW · Taiwan

Affected area 55

UA · Ukraine

Affected area 56

UY · Uruguay

Affected area 57

VN · Vietnam

Affected area 58

ZA · South Africa

Company & contacts

Manufacturer

Draeger Medical Systems, Inc.

Phone

Not provided

Brand: Draeger Medical. Contact: Draeger Medical Systems, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2189-2024

Event ID

94644

Units affected

34,102

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 3 Jul 2024. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button