Open official notice

Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833

Official notice

16 Apr 2014

RecallScope risk estimate:CriticalFDA

Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

3

Recall number

Z-1440-2014

Official notice

2014-04-16

The hazard

The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

67677

Recall Numbers

Z-1440-2014

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833

Units affected

3

Hazard

The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Draeger Medical Systems, Inc.

Phone

Not provided

Brand: Draeger Medical. Contact: Draeger Medical Systems, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class I

Recall details

Recall number

Z-1440-2014

Event ID

67677

Units affected

3

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 16 Apr 2014. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.