Image not available in RecallScope.
Babytherm 8004 8010, Infant Warmer, Class II device, 510 K K971198, Catalog # FR00105 (Babytherm 8010), 2M30404 (Babytherm 8004), non
Official notice
3 Aug 2016
Babytherm 8004/8010, Infant Warmer, Class II device, 510 K K971198, Catalog # FR00105 (Babytherm 8010), 2M30404 (Babytherm 8004), non sterile, software controlled. Product Usage: Babytherm 8004/8010 is an open care infant warmer system for warming premature babies, neonates and infants with a body weight of up to 8 kg (17.6 lbs). The device is intended for use in delivery rooms, operating rooms, maternity and obstetric units, neonatal and pediatric wards, neonatal and pediatric intensive care units. Therapy and nursing uses: - As a resuscitation table and in birthing rooms, obstetric units, and NICUs - As infant warmer bassinet during normal and intensive care - For infant re-warming and compensation of heat loss - For thermal stabilizing of infants with mattress heater (Babytherm 8010 only) - For gentle warming therapy using the combination of radiant and mattress heater - Cooling infants with fever (temperature of mattress below infant body temperature, Babytherm 8010 only) - With adjustable bed tilt for treatment of infants in Trendelenburg and reverse Trendelenburg position - For weaning infants from incubators - For O2 and nebulizer therapy (with appropriate accessories) - For lowering an infant s bilirubin levels when using the phototherapy option
What to do now
This recall has been terminated. See the official FDA notice for the latest status.
Units affected
1,599
Recall number
Z-2295-2016
Official notice
2016-08-03
The hazard
The hazard
Risk associated with operating the device in skin temperature mode with temperature sensors attached but no patient connected.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Draeger Medical. Contact: Draeger Medical, Inc..
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source