Eon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a

Official notice

31 Jul 2013

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Eon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a. There is issues of warmth or heating at the implant site during charging for the Eon Mini implantable pulse generators (IPGs). In the July 2012 letter St. Jude Medical informed customers that they would be implementing design improvements to the charger to address possible increased energy dissipation when the charger is misaligned with the IPG or the IPG is implanted too near the surface of the Published 2013-07-31.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

75,663

Recall number

Z-1785-2013

Official notice

2013-07-31

The hazard

There is issues of warmth or heating at the implant site during charging for the Eon Mini implantable pulse generators (IPGs). In the July 2012 letter St. Jude Medical informed customers that they would be implementing design improvements to the charger to address possible increased energy dissipation when the charger is misaligned with the IPG or the IPG is implanted too near the surface of the

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

63947

Recall Numbers

Z-1785-2013

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Eon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 recall. Product Usage: The Eon Wall Charger (3701) provides the capability to recharge the IPG Battery. The charger transmits RF energy through the antenna to the IPG battery to recharge it. The Eon Portable Charger (3711) provides the capability to recharge the IPG battery. The charger transmits RF energy through the antenna to the IPG battery to recharge it. The patient can be mobile during the recharging process with a portable charger. The Eon Mini Charger (3721) provides the capability to recharge the IPG battery while stimulation is either on or off. The charger transmits RF energy through the antenna to the IPG battery to recharge it.

Units affected

75,663

Hazard

There is issues of warmth or heating at the implant site during charging for the Eon Mini implantable pulse generators (IPGs). In the July 2012 letter St. Jude Medical informed customers that they would be implementing design improvements to the charger to address possible increased energy dissipation when the charger is misaligned with the IPG or the IPG is implanted too near the surface of the

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AT · Austria

Affected area 3

AU · Australia

Affected area 4

BE · Belgium

Affected area 5

BR · Brazil

Affected area 6

CA · Canada

Affected area 7

CH · Switzerland

Affected area 8

CL · Chile

Affected area 9

CN · China

Affected area 10

CO · Colombia

Affected area 11

CY · Cyprus

Affected area 12

CZ · Czech Republic

Affected area 13

DE · Germany

Affected area 14

DK · Denmark

Affected area 15

ES · Spain

Affected area 16

FI · Finland

Affected area 17

FR · France

Affected area 18

GB · United Kingdom

Affected area 19

GR · Greece

Affected area 20

HK · Hong Kong

Affected area 21

IE · Ireland

Affected area 22

IL · Israel

Affected area 23

IN · India

Affected area 24

IT · Italy

Affected area 25

JP · Japan

Affected area 26

KR · South Korea

Affected area 27

LU · Luxembourg

Affected area 28

MX · Mexico

Affected area 29

NL · Netherlands

Affected area 30

PL · Poland

Affected area 31

SA · Saudi Arabia

Affected area 32

SE · Sweden

Affected area 33

SG · Singapore

Affected area 34

TR · Turkey

Affected area 35

ZA · South Africa

Company & contacts

Manufacturer

St. Jude Medical

Phone

Not provided

Brand: St Jude Medical. Contact: St. Jude Medical.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1785-2013

Event ID

63947

Units affected

75,663

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 31 Jul 2013. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button