Open official notice

Ellipse vr, ref: cd1377-36c implantable cardioverter defibrillators (icds)

Official notice

14 Aug 2019

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns Ellipse vr, ref: cd1377-36c implantable cardioverter defibrillators (icds). Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices. Published 2019-08-14.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

27

Recall number

Z-2070-2019

Official notice

2019-08-14

The hazard

Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

83252

Recall Numbers

Z-2070-2019

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

ELLIPSE VR, REF: CD1377-36C Implantable cardioverter defibrillators (ICDs)

Units affected

27

Hazard

Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

St Jude Medical Inc.

Phone

Not provided

Brand: St Jude Medical. Contact: St Jude Medical Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class I

Recall details

Recall number

Z-2070-2019

Event ID

83252

Units affected

27

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 14 Aug 2019. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.