Open official notice

St Jude Medical, 10F Fast-Cath Trio Hemostasis Introducer, REF # 406308, Sterile EO. The Fast-Cath Trio Hemostasis Introducer consists of

Official notice

29 Oct 2014

RecallScope risk estimate:MediumFDA

An official FDA notice concerns St Jude Medical, 10F Fast-Cath Trio Hemostasis Introducer, REF # 406308, Sterile EO. The Fast-Cath Trio Hemostasis Introducer consists of. St. Jude Medical is conducting a voluntary recall of SMJ Batch # 4641580 of the Fast-Cath Trio" Hemostasis Introducer, Catalog number 406308. The inner diameter of the 10F dilator is too small, such that the guidewire cannot advance through the device which may result in a procedural delay while another dilator of the desired size is obtained. Published 2014-10-29.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

220

Recall number

Z-0108-2015

Official notice

2014-10-29

The hazard

St. Jude Medical is conducting a voluntary recall of SMJ Batch # 4641580 of the Fast-Cath Trio" Hemostasis Introducer, Catalog number 406308. The inner diameter of the 10F dilator is too small, such that the guidewire cannot advance through the device which may result in a procedural delay while another dilator of the desired size is obtained.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

4641580

Event Ids

69499

Recall Numbers

Z-0108-2015

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

St Jude Medical, 10F Fast-Cath Trio" Hemostasis Introducer, REF # 406308, Sterile EO. The Fast-Cath Trio" Hemostasis Introducer consists of the following components: Port hemostasis adapter (3), Sheath (10F Cath-Lock"), dilator (10F) and 0.038 guidewire. See product labeling for diagram of contents.

Units affected

220

Hazard

St. Jude Medical is conducting a voluntary recall of SMJ Batch # 4641580 of the Fast-Cath Trio" Hemostasis Introducer, Catalog number 406308. The inner diameter of the 10F dilator is too small, such that the guidewire cannot advance through the device which may result in a procedural delay while another dilator of the desired size is obtained.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

St Jude Medical

Phone

Not provided

Brand: St Jude Medical. Contact: St Jude Medical.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0108-2015

Event ID

69499

Units affected

220

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 29 Oct 2014. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button