DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers UDI codes: C63519 14987666545058 C63520 14987666545065. The Beckman Coulter DxC

Official notice

21 May 2025

RecallScope risk estimate:MediumFDA

An official FDA notice concerns DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers UDI codes: C63519 14987666545058 C63520 14987666545065. The Beckman Coulter DxC. Beckman Coulter is recalling their DxC 500 AU Clinical Chemistry Analyzer because a software error causes the analyzer to not run a requested calibration order in the following scenario: when a reagent blank or calibration is ordered during sample processing and then any of the components (R1 and/or R2) depletes to zero tests, the analyzer will not be able to complete the calibration request, and the calibration order will remain pending. No further calibration orders can be processed for any assays, and the instrument refuses to accept further sample processing order after the existing calibration curves are expired. Although in-process tests will be completed, this error can cause a delay in reporting subsequent test results. No further calibration orders can be processed for any assays which may cause a delay in reporting test results. Published 2025-05-21.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Recall number

Z-1784-2025

Official notice

2025-05-21

The hazard

Beckman Coulter is recalling their DxC 500 AU Clinical Chemistry Analyzer because a software error causes the analyzer to not run a requested calibration order in the following scenario: when a reagent blank or calibration is ordered during sample processing and then any of the components (R1 and/or R2) depletes to zero tests, the analyzer will not be able to complete the calibration request, and the calibration order will remain pending. No further calibration orders can be processed for any assays, and the instrument refuses to accept further sample processing order after the existing calibration curves are expired. Although in-process tests will be completed, this error can cause a delay in reporting subsequent test results. No further calibration orders can be processed for any assays which may cause a delay in reporting test results.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

UDI-DI 14987666545058, UDI-DI 14987666545065

Event Ids

96512

Recall Numbers

Z-1784-2025

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065. The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.

Hazard

Beckman Coulter is recalling their DxC 500 AU Clinical Chemistry Analyzer because a software error causes the analyzer to not run a requested calibration order in the following scenario: when a reagent blank or calibration is ordered during sample processing and then any of the components (R1 and/or R2) depletes to zero tests, the analyzer will not be able to complete the calibration request, and the calibration order will remain pending. No further calibration orders can be processed for any assays, and the instrument refuses to accept further sample processing order after the existing calibration curves are expired. Although in-process tests will be completed, this error can cause a delay in reporting subsequent test results. No further calibration orders can be processed for any assays which may cause a delay in reporting test results.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AL

Location 2

US · United States · CA

Location 3

US · United States · GA

Location 4

US · United States · ID

Location 5

US · United States · IN

Location 6

US · United States · LA

Location 7

US · United States · MA

Location 8

US · United States · MI

Location 9

US · United States · MN

Location 10

US · United States · MT

Location 11

US · United States · NC

Location 12

US · United States · NV

Location 13

US · United States · PA

Location 14

US · United States · TN

Location 15

US · United States · TX

Location 16

US · United States · WI

Affected area 1

US · United States

Affected area 2

AU · Australia

Affected area 3

BE · Belgium

Affected area 4

CA · Canada

Affected area 5

CH · Switzerland

Affected area 6

CL · Chile

Affected area 7

DE · Germany

Affected area 8

EG · Egypt

Affected area 9

ES · Spain

Affected area 10

FR · France

Affected area 11

HR · Croatia

Affected area 12

IN · India

Affected area 13

IT · Italy

Affected area 14

KR · South Korea

Affected area 15

MY · Malaysia

Affected area 16

NL · Netherlands

Affected area 17

PL · Poland

Affected area 18

SA · Saudi Arabia

Affected area 19

ZA · South Africa

Company & contacts

Manufacturer

Beckman Coulter Inc.

Phone

Not provided

Brand: Beckman Coulter. Contact: Beckman Coulter Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1784-2025

Event ID

96512

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 21 May 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button