Open official notice

Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; Catalog Number: (1) L2KPP2, (2)

Official notice

5 Nov 2025

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; Catalog Number: (1) L2KPP2, (2). The potential for falsely depressed Intact PTH patient results at the low end of the assay range, less than or equal to 50 pg/mL (less than or equal to 5.3 pmol/L), when using specific lots identified by the firm on the IMMULITE 2000/IMMULITE 2000 XPi systems. The bias is observed in both serum and plasma samples. Published 2025-11-05.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

1,740

Recall number

Z-0366-2026

Official notice

2025-11-05

The hazard

The potential for falsely depressed Intact PTH patient results at the low end of the assay range, less than or equal to 50 pg/mL (less than or equal to 5.3 pmol/L), when using specific lots identified by the firm on the IMMULITE 2000/IMMULITE 2000 XPi systems. The bias is observed in both serum and plasma samples.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

UDI-DI 00630414961910, UDI-DI 00630414961927

Event Ids

97672

Recall Numbers

Z-0366-2026

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; Catalog Number: (1) L2KPP2, (2) L2KPP6

Units affected

1,740

Hazard

The potential for falsely depressed Intact PTH patient results at the low end of the assay range, less than or equal to 50 pg/mL (less than or equal to 5.3 pmol/L), when using specific lots identified by the firm on the IMMULITE 2000/IMMULITE 2000 XPi systems. The bias is observed in both serum and plasma samples.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AR · Argentina

Affected area 3

AU · Australia

Affected area 4

BG · Bulgaria

Affected area 5

BR · Brazil

Affected area 6

CA · Canada

Affected area 7

CL · Chile

Affected area 8

CN · China

Affected area 9

CO · Colombia

Affected area 10

CY · Cyprus

Affected area 11

CZ · Czech Republic

Affected area 12

DE · Germany

Affected area 13

EG · Egypt

Affected area 14

ES · Spain

Affected area 15

FR · France

Affected area 16

GR · Greece

Affected area 17

GT · Guatemala

Affected area 18

HU · Hungary

Affected area 19

IL · Israel

Affected area 20

IN · India

Affected area 21

IT · Italy

Affected area 22

KW · Kuwait

Affected area 23

LV · Latvia

Affected area 24

MX · Mexico

Affected area 25

NL · Netherlands

Affected area 26

PE · Peru

Affected area 27

PK · Pakistan

Affected area 28

PL · Poland

Affected area 29

PT · Portugal

Affected area 30

RO · Romania

Affected area 31

RS · Serbia

Affected area 32

SI · Slovenia

Affected area 33

SK · Slovakia

Affected area 34

TR · Turkey

Affected area 35

TW · Taiwan

Affected area 36

UA · Ukraine

Affected area 37

UY · Uruguay

Affected area 38

ZA · South Africa

Company & contacts

Brand

Manufacturer

Siemens Healthcare Diagnostics, Inc.

Phone

Not provided

Brand: Siemens. Contact: Siemens Healthcare Diagnostics, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0366-2026

Event ID

97672

Units affected

1,740

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 5 Nov 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button