Open official notice

Product: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300

Official notice

2 Jul 2025

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Product: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300. Contamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride. Published 2025-07-02.

What to do now

Do not consume or use the affected product. Return it to the place of purchase or follow instructions in the official FDA notice.

Units affected

1,625

Recall number

Z-2027-2025

Official notice

2025-07-02

The hazard

Contamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

0000195412

Other

UDI-DI 00630414595771

Event Ids

96952

Recall Numbers

Z-2027-2025

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Product: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300

Units affected

1,625

Hazard

Contamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

CA · Canada

Company & contacts

Brand

Manufacturer

Siemens Healthcare Diagnostics, Inc.

Phone

Not provided

Brand: Siemens. Contact: Siemens Healthcare Diagnostics, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2027-2025

Event ID

96952

Units affected

1,625

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 2 Jul 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button