Epoc BGEM BUN Test Card [25pk]. Material Number: 10736515

Official notice

13 Aug 2025

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Epoc BGEM BUN Test Card [25pk]. Material Number: 10736515. Siemens Healthcare Diagnostics Inc. Has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L. The maximum bias observed was -14 mmol/L, which occurred at a higher sodium concentration around 150 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. Published 2025-08-13.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

37,554

Recall number

Z-2272-2025

Official notice

2025-08-13

The hazard

Siemens Healthcare Diagnostics Inc. Has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L. The maximum bias observed was -14 mmol/L, which occurred at a higher sodium concentration around 150 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

97180

Recall Numbers

Z-2272-2025

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

epoc BGEM BUN Test Card [25pk]. Material Number: 10736515.

Units affected

37,554

Hazard

Siemens Healthcare Diagnostics Inc. Has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L. The maximum bias observed was -14 mmol/L, which occurred at a higher sodium concentration around 150 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AR · Argentina

Affected area 3

AT · Austria

Affected area 4

BR · Brazil

Affected area 5

CA · Canada

Affected area 6

CH · Switzerland

Affected area 7

CL · Chile

Affected area 8

DE · Germany

Affected area 9

DK · Denmark

Affected area 10

EE · Estonia

Affected area 11

ES · Spain

Affected area 12

FR · France

Affected area 13

GB · United Kingdom

Affected area 14

GR · Greece

Affected area 15

GT · Guatemala

Affected area 16

HR · Croatia

Affected area 17

HU · Hungary

Affected area 18

IE · Ireland

Affected area 19

IL · Israel

Affected area 20

IN · India

Affected area 21

IS · Iceland

Affected area 22

IT · Italy

Affected area 23

JO · Jordan

Affected area 24

JP · Japan

Affected area 25

KR · South Korea

Affected area 26

LT · Lithuania

Affected area 27

MX · Mexico

Affected area 28

MY · Malaysia

Affected area 29

NL · Netherlands

Affected area 30

PL · Poland

Affected area 31

PT · Portugal

Affected area 32

RO · Romania

Affected area 33

SE · Sweden

Affected area 34

SI · Slovenia

Affected area 35

SK · Slovakia

Affected area 36

TH · Thailand

Affected area 37

TR · Turkey

Affected area 38

TW · Taiwan

Affected area 39

UA · Ukraine

Affected area 40

UY · Uruguay

Affected area 41

ZA · South Africa

Company & contacts

Brand

Manufacturer

Siemens Healthcare Diagnostics Inc

Phone

Not provided

Brand: Siemens. Contact: Siemens Healthcare Diagnostics Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2272-2025

Event ID

97180

Units affected

37,554

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 13 Aug 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button