VITEK 2 Gram Positive Susceptibility Test Cards (AST-P605), REF 22 325. 20 Cards Kit. BioMerieux, Inc, Durham, North Carolina 27712. VITEK

Official notice

11 May 2016

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns VITEK 2 Gram Positive Susceptibility Test Cards (AST-P605), REF 22 325. 20 Cards Kit. BioMerieux, Inc, Durham, North Carolina 27712. VITEK. Product may lead to false susceptible results of Staphylococcus spp., Enterococcus spp., and S. Agalactiae to Minocycline. Published 2016-05-11.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

337

Recall number

Z-1596-2016

Official notice

2016-05-11

The hazard

Product may lead to false susceptible results of Staphylococcus spp., Enterococcus spp., and S. Agalactiae to Minocycline.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

73585

Recall Numbers

Z-1596-2016

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

VITEK 2 Gram Positive Susceptibility Test Cards (AST-P605), REF 22 325. 20 Cards/Kit. bioMerieux, Inc, Durham, North Carolina 27712. VITEK 2 is an automated system consisting of instruments, software, and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. The VITEK 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria. The VITEK 2 System is not considered a life supporting/sustaining device and does not deliver any type of energy or substance to the patient. The VITEK 2 Gram Positive Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed in the Product Information manual.

Units affected

337

Hazard

Product may lead to false susceptible results of Staphylococcus spp., Enterococcus spp., and S. Agalactiae to Minocycline.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

CN · China

Affected area 3

TW · Taiwan

Company & contacts

Manufacturer

Biomerieux Inc

Phone

Not provided

Brand: Biomerieux. Contact: Biomerieux Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1596-2016

Event ID

73585

Units affected

337

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 11 May 2016. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button