VIDAS D-Dimer Exclusion II, an automated quantitative test for use on instruments of the VIDAS family for the immunoenzymatic determination
Official notice
20 Aug 2014
An official FDA notice concerns VIDAS D-Dimer Exclusion II, an automated quantitative test for use on instruments of the VIDAS family for the immunoenzymatic determination. Customer complaints about not repeatable results below the detection limit (<45 ng/mL). Therefore false negatives are being reported for results around the medical decision point of 500 ng/mL that initially tested at <45 ng/mL. Published 2014-08-20.
What to do now
This recall has been terminated. See the official FDA notice for the latest status.
Units affected
2,160
Recall number
Z-2232-2014
Official notice
2014-08-20
The hazard
The hazard
Customer complaints about not repeatable results below the detection limit (<45 ng/mL). Therefore false negatives are being reported for results around the medical decision point of 500 ng/mL that initially tested at <45 ng/mL.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Location 2
Location 3
Location 4
Location 5
Location 6
Location 7
Location 8
Location 9
Location 10
Location 11
Location 12
Location 13
Location 14
Location 15
Location 16
Location 17
Location 18
Location 19
Location 20
Location 21
Location 22
Location 23
Location 24
Location 25
Location 26
Location 27
Location 28
Location 29
Location 30
Location 31
Location 32
Location 33
Location 34
Location 35
Location 36
Location 37
Location 38
Location 39
Location 40
Affected area 1
Company & contacts
Company & contacts
Brand: Biomerieux. Contact: BioMerieux SA.
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.