Open official notice

Tosoh Bioscience HLC-723G8 Analyzer for HbA1c testing

Official notice

1 Mar 2017

RecallScope risk estimate:MediumFDA

Tosoh Bioscience HLC-723G8 Analyzer for HbA1c testing.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

682

Recall number

Z-1225-2017

Official notice

2017-03-01

The hazard

The device defect associated with this recall may result in HbA1c values with a negative bias in patients with the hemoglobin variants C, D, E and S. The negative bias occurs at the higher levels of HbA1c, e.g. 9%, but not at the diagnostic medical decision point of 6.5%.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

75727

Recall Numbers

Z-1225-2017

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Tosoh Bioscience HLC-723G8 Analyzer for HbA1c testing.

Units affected

682

Hazard

The device defect associated with this recall may result in HbA1c values with a negative bias in patients with the hemoglobin variants C, D, E and S. The negative bias occurs at the higher levels of HbA1c, e.g. 9%, but not at the diagnostic medical decision point of 6.5%.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AK

Location 2

US · United States · AL

Location 3

US · United States · AR

Location 4

US · United States · AZ

Location 5

US · United States · CA

Location 6

US · United States · CO

Location 7

US · United States · CT

Location 8

US · United States · DE

Location 9

US · United States · FL

Location 10

US · United States · GA

Location 11

US · United States · IA

Location 12

US · United States · ID

Location 13

US · United States · IL

Location 14

US · United States · IN

Location 15

US · United States · KS

Location 16

US · United States · KY

Location 17

US · United States · LA

Location 18

US · United States · MA

Location 19

US · United States · MD

Location 20

US · United States · ME

Location 21

US · United States · MI

Location 22

US · United States · MN

Location 23

US · United States · MO

Location 24

US · United States · MS

Location 25

US · United States · NC

Location 26

US · United States · ND

Location 27

US · United States · NE

Location 28

US · United States · NH

Location 29

US · United States · NJ

Location 30

US · United States · NM

Location 31

US · United States · NV

Location 32

US · United States · NY

Location 33

US · United States · OH

Location 34

US · United States · OK

Location 35

US · United States · OR

Location 36

US · United States · PA

Location 37

US · United States · RI

Location 38

US · United States · SC

Location 39

US · United States · SD

Location 40

US · United States · TN

Location 41

US · United States · TX

Location 42

US · United States · UT

Location 43

US · United States · VA

Location 44

US · United States · VT

Location 45

US · United States · WA

Location 46

US · United States · WI

Location 47

US · United States · WV

Location 48

US · United States · WY

Affected area 1

US · United States

Affected area 2

BR · Brazil

Affected area 3

CL · Chile

Affected area 4

CR · Costa Rica

Affected area 5

DO · Dominican Republic

Affected area 6

GT · Guatemala

Affected area 7

PE · Peru

Affected area 8

UY · Uruguay

Company & contacts

Manufacturer

Tosoh Bioscience, Inc.

Phone

Not provided

Brand: Tosoh Bioscience. Contact: Tosoh Bioscience, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1225-2017

Event ID

75727

Units affected

682

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 1 Mar 2017. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button