AIA-900 Automated Immunoassay Analyzer, product code 022930 022930R Product Usage: The AIA-900 is capable of performing three methods of

Official notice

5 Dec 2018

RecallScope risk estimate:MediumFDA

AIA-900 Automated Immunoassay Analyzer, product code 022930 /022930R Product Usage: The AIA-900 is capable of performing three methods of immunoassay: an immunoenzymetric (IEMA) or sandwich immunoassay, a competitive binding (EIA) immunoassay, and a two-step immunoenzymetric immunoassay. An antigen-antibody reaction begins by combining a patient sample, control, or calibrator with a diluent in an immunoreaction test cup from the AIA-PACK reagent series.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

247

Recall number

Z-0509-2019

Official notice

2018-12-05

The hazard

Tosoh Bioscience has become aware of a potential issue with the axis base of the turntable rotation drive motor on the AIA-900 Analyzer. If the tension of the timing belt between the turntable rotation pulley and drive motor is not adjusted correctly, it is possible to increase the lateral load on the motor axis resulting in the inclination and damage of the motor axis. Since January 12, 2017, Tosoh has been inspecting the tension of the timing belt during scheduled preventative maintenance visit and correcting the analyzers where this issue has been observed. If the drive motor fails, the AIA-900 analyzer will cease to function and will be unavailable for use. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh has received five (5) complaints related to this issue with no serious injuries reported.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

81212

Recall Numbers

Z-0509-2019

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

AIA-900 Automated Immunoassay Analyzer, product code 022930 /022930R Product Usage: The AIA-900 is capable of performing three methods of immunoassay: an immunoenzymetric (IEMA) or sandwich immunoassay, a competitive binding (EIA) immunoassay, and a two-step immunoenzymetric immunoassay. An antigen-antibody reaction begins by combining a patient sample, control, or calibrator with a diluent in an immunoreaction test cup from the AIA-PACK reagent series.

Units affected

247

Hazard

Tosoh Bioscience has become aware of a potential issue with the axis base of the turntable rotation drive motor on the AIA-900 Analyzer. If the tension of the timing belt between the turntable rotation pulley and drive motor is not adjusted correctly, it is possible to increase the lateral load on the motor axis resulting in the inclination and damage of the motor axis. Since January 12, 2017, Tosoh has been inspecting the tension of the timing belt during scheduled preventative maintenance visit and correcting the analyzers where this issue has been observed. If the drive motor fails, the AIA-900 analyzer will cease to function and will be unavailable for use. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh has received five (5) complaints related to this issue with no serious injuries reported.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

CL · Chile

Affected area 3

DO · Dominican Republic

Affected area 4

GT · Guatemala

Affected area 5

PE · Peru

Affected area 6

UY · Uruguay

Affected area 7

VE · Venezuela

Company & contacts

Manufacturer

Tosoh Bioscience Inc

Phone

Not provided

Brand: Tosoh Bioscience. Contact: Tosoh Bioscience Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0509-2019

Event ID

81212

Units affected

247

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 5 Dec 2018. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button