Open official notice

The Medline General Surgery Tray is customized to meet requirements of the hospitals and health care professionals utilizing them. The kit

Official notice

3 Sept 2025

RecallScope risk estimate:MediumFDA

The Medline General Surgery Tray is customized to meet requirements of the hospitals and health care professionals utilizing them. The kit is provided sterile, single-use and includes a range of components dependent upon the specific procedure. These kits will be assembled using Medline and other manufacturer s medical device components and drug components in some cases. Customers specify the kit contents, quantity, and placement of individual items in the kit.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

1,145

Recall number

Z-2452-2025

Official notice

2025-09-03

The hazard

Affected kits contain recalled Stryker Color Cuff Sterile Disposable Tourniquet Cuffs. Stryker Instruments initiated this recall due to a potential manufacturing defect in which the flange may detach from the bladder at the welded connection point. This detachment compromises the cuff's ability to maintain adequate air pressure, potentially resulting in tourniquet failure and increased risk of patient hemorrhaging.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

25CMH791

Event Ids

97371

Recall Numbers

Z-2452-2025

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

The Medline General Surgery Tray is customized to meet requirements of the hospitals and health care professionals utilizing them. The kit is provided sterile, single-use and includes a range of components dependent upon the specific procedure. These kits will be assembled using Medline and other manufacturers medical device components and drug components in some cases. Customers specify the kit contents, quantity, and placement of individual items in the kit.

Units affected

1,145

Hazard

Affected kits contain recalled Stryker Color Cuff Sterile Disposable Tourniquet Cuffs. Stryker Instruments initiated this recall due to a potential manufacturing defect in which the flange may detach from the bladder at the welded connection point. This detachment compromises the cuff's ability to maintain adequate air pressure, potentially resulting in tourniquet failure and increased risk of patient hemorrhaging.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

MEDLINE INDUSTRIES, LP - Northfield

Phone

Not provided

Brand: Medline Lp - Northfield. Contact: MEDLINE INDUSTRIES, LP - Northfield.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2452-2025

Event ID

97371

Units affected

1,145

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 3 Sept 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button