Open official notice

SwabFlush Prefilled Syringe with ICU Medical SwabCaps, 10mL Model no. EMZE010301

Official notice

6 Aug 2025

RecallScope risk estimate:MediumFDA

SwabFlush Prefilled Syringe with ICU Medical SwabCaps, 10mL Model no. EMZE010301

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

1,427,100

Recall number

Z-2199-2025

Official notice

2025-08-06

The hazard

SwabFlush syringes are affected by the ICU Medical recall of their SwabCap Disinfecting Cap due to a manufacturing issue which affects the seal between the foil lid and plastic container of the Swab Cap. The issue may result in a reduced level of disinfection of the luer access valve, potentially resulting in patient infection/sepsis.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

UDI-DI 00363807103016

Event Ids

97169

Recall Numbers

Z-2199-2025

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

SwabFlush Prefilled Syringe with ICU Medical SwabCaps, 10mL Model no. EMZE010301

Units affected

1,427,100

Hazard

SwabFlush syringes are affected by the ICU Medical recall of their SwabCap Disinfecting Cap due to a manufacturing issue which affects the seal between the foil lid and plastic container of the Swab Cap. The issue may result in a reduced level of disinfection of the luer access valve, potentially resulting in patient infection/sepsis.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

CA · Canada

Company & contacts

Manufacturer

MEDLINE INDUSTRIES, LP - Northfield

Phone

Not provided

Brand: Medline Lp - Northfield. Contact: MEDLINE INDUSTRIES, LP - Northfield.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2199-2025

Event ID

97169

Units affected

1,427,100

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 6 Aug 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button