Open official notice

Safety monitor included with System 8000 Base The Sarns Modular Perfusion System 8000 is indicated for use for up to 6 hours on the

Official notice

19 Sept 2012

RecallScope risk estimate:MediumFDA

Safety monitor included with System 8000 Base The Sarns" Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. The safety monitor is indicated to warn when the blood level in the reservoir reaches the operator-set limit, and to stop the arterial and cardioplegia pumps.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

8

Recall number

Z-2348-2012

Official notice

2012-09-19

The hazard

Terumo Cardiovascular Systems (Terumo CVS) received complaints from our Affiliate, Terumo Dubai that the safety monitors were tripping off. Terumo CVS' investigation determined that 8 monitors were manufactured with an underrated circuit breaker due to an error where an associate pulled an underrated circuit breaker from an incorrect inventory location. Since this time, inventory locations hav

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

62321

Recall Numbers

Z-2348-2012

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Safety monitor included with System 8000 Base The Sarns" Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. The safety monitor is indicated to warn when the blood level in the reservoir reaches the operator-set limit, and to stop the arterial and cardioplegia pumps.

Units affected

8

Hazard

Terumo Cardiovascular Systems (Terumo CVS) received complaints from our Affiliate, Terumo Dubai that the safety monitors were tripping off. Terumo CVS' investigation determined that 8 monitors were manufactured with an underrated circuit breaker due to an error where an associate pulled an underrated circuit breaker from an incorrect inventory location. Since this time, inventory locations hav

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Company & contacts

Manufacturer

Terumo Cardiovascular Systems Corporation

Phone

Not provided

Brand: Terumo Cardiovascular. Contact: Terumo Cardiovascular Systems Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2348-2012

Event ID

62321

Units affected

8

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 19 Sept 2012. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button