Open official notice

Hercules 360 Universal Stabilizing Arm, top mount, 360-degree rotation, reusable stainless steel

Official notice

22 Feb 2017

RecallScope risk estimate:MediumFDA

Hercules 360 Universal Stabilizing Arm, top mount, 360-degree rotation, reusable stainless steel

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

260

Recall number

Z-1201-2017

Official notice

2017-02-22

The hazard

Terumo Cardiovascular Systems (Terumo CVS) is recalling certain lots of Hercules TM 360 Universal Stabilizing Arms because the two pins that hold the sleeve attached to the distal end of the device may come loose, causing the pins and sleeve to separate from the device. While the arm can still function without the pins and sleeve in place, there is potential for the loose components to fall into the surgical site.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

76344

Recall Numbers

Z-1201-2017

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Hercules 360 Universal Stabilizing Arm, top mount, 360-degree rotation, reusable stainless steel

Units affected

260

Hazard

Terumo Cardiovascular Systems (Terumo CVS) is recalling certain lots of Hercules TM 360 Universal Stabilizing Arms because the two pins that hold the sleeve attached to the distal end of the device may come loose, causing the pins and sleeve to separate from the device. While the arm can still function without the pins and sleeve in place, there is potential for the loose components to fall into the surgical site.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Terumo Cardiovascular Systems Corporation

Phone

Not provided

Brand: Terumo Cardiovascular. Contact: Terumo Cardiovascular Systems Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1201-2017

Event ID

76344

Units affected

260

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 22 Feb 2017. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button