Open official notice

Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm

Official notice

12 Mar 2025

RecallScope risk estimate:CriticalFDA

Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm, REF: 80238; S 8 F x 110 cm, REF: 80239

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

15,755

Recall number

Z-1244-2025

Official notice

2025-03-12

The hazard

Atherectomy Catheter eIFU updated to clarify and emphasize procedural steps intended to reduce the likelihood of catheter breakage events. Catheter has an outer cylinder connected to a rotating helix, which could fracture and/or break, which would require retrieval, and helix facture/break could cause vessel injury and lead to severe bleeding.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

96174

Recall Numbers

Z-1244-2025

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm, REF: 80238; S 8 F x 110 cm, REF: 80239

Units affected

15,755

Hazard

Atherectomy Catheter eIFU updated to clarify and emphasize procedural steps intended to reduce the likelihood of catheter breakage events. Catheter has an outer cylinder connected to a rotating helix, which could fracture and/or break, which would require retrieval, and helix facture/break could cause vessel injury and lead to severe bleeding.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Bard Peripheral Vascular Inc

Phone

Not provided

Brand: Bard Peripheral Vascular. Contact: Bard Peripheral Vascular Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class I

Recall details

Recall number

Z-1244-2025

Event ID

96174

Units affected

15,755

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 12 Mar 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button