Open official notice

Bard Mission Disposable Core Biopsy Instrument Kit (REF Name): 1410MSK Mission Kit 14G x 10CM; 1416MSK Mission Kit 14G x 16CM; 1610MSK

Official notice

17 Jan 2024

RecallScope risk estimate:CriticalFDA

Bard Mission Disposable Core Biopsy Instrument Kit (REF/Name): 1410MSK/ Mission Kit 14G x 10CM; 1416MSK/ Mission Kit 14G x 16CM; 1610MSK/ Mission Kit 16G x 10CM; 1616MSK/ Mission Kit 16G x 16CM; 1810MSK/ Mission Kit 18G x 10CM; 1816MSK/ Mission Kit 18G x 16CM; 1820MSK/ Mission Kit 18G x 20CM; 1825MSK/ Mission Kit 18G x 25CM; 2010MSK/ Mission Kit 20G x 10CM; 2016MSK/ Mission Kit 20G x 16CM; 2020MSK/ Mission Kit 20G x 20CM

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

1,243,937

Recall number

Z-0643-2024

Official notice

2024-01-17

The hazard

The internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy needle may not fit properly into the coaxial cannula, preventing access to the target tissue, which could lead to tissue injury, longer and repeat procedures.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

93578

Recall Numbers

Z-0643-2024

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Bard Mission Disposable Core Biopsy Instrument Kit (REF/Name): 1410MSK/ Mission Kit 14G x 10CM; 1416MSK/ Mission Kit 14G x 16CM; 1610MSK/ Mission Kit 16G x 10CM; 1616MSK/ Mission Kit 16G x 16CM; 1810MSK/ Mission Kit 18G x 10CM; 1816MSK/ Mission Kit 18G x 16CM; 1820MSK/ Mission Kit 18G x 20CM; 1825MSK/ Mission Kit 18G x 25CM; 2010MSK/ Mission Kit 20G x 10CM; 2016MSK/ Mission Kit 20G x 16CM; 2020MSK/ Mission Kit 20G x 20CM

Units affected

1,243,937

Hazard

The internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy needle may not fit properly into the coaxial cannula, preventing access to the target tissue, which could lead to tissue injury, longer and repeat procedures.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Bard Peripheral Vascular Inc

Phone

Not provided

Brand: Bard Peripheral Vascular. Contact: Bard Peripheral Vascular Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0643-2024

Event ID

93578

Units affected

1,243,937

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 17 Jan 2024. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button