Paradigm REF: MMT-712 and MMT-715
Official notice
12 Mar 2025
An official FDA notice concerns Paradigm REF: MMT-712 and MMT-715. All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. During a delivery accuracy dry run testing of the 780G NGP infusion pump under dynamic pressure conditions, the firm observed that when air pressure decreased during take-off simulation, on average 0.569U more insulin was delivered in 15 minutes than the expected delivery volume at ambient pressure at 1 U/hr delivery rate; when air pressure increased during landing simulation, on average 0.537 U less insulin was delivered in 25 minutes than the expected delivery volume at ambient pressure at 1 U/hr delivery rate. The firm also observed that the mechanisms causing the variations in delivery volumes are independent of the delivery rate. Published 2025-03-12.
What to do now
Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.
Units affected
20,760
Recall number
Z-1255-2025
Official notice
2025-03-12
The hazard
The hazard
All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. During a delivery accuracy dry run testing of the 780G NGP infusion pump under dynamic pressure conditions, the firm observed that when air pressure decreased during take-off simulation, on average 0.569U more insulin was delivered in 15 minutes than the expected delivery volume at ambient pressure at 1 U/hr delivery rate; when air pressure increased during landing simulation, on average 0.537 U less insulin was delivered in 25 minutes than the expected delivery volume at ambient pressure at 1 U/hr delivery rate. The firm also observed that the mechanisms causing the variations in delivery volumes are independent of the delivery rate.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Affected area 2
Affected area 3
Affected area 4
Affected area 5
Affected area 6
Affected area 7
Affected area 8
Affected area 9
Affected area 10
Affected area 11
Affected area 12
Affected area 13
Affected area 14
Affected area 15
Affected area 16
Affected area 17
Affected area 18
Affected area 19
Affected area 20
Affected area 21
Affected area 22
Affected area 23
Affected area 24
Affected area 25
Affected area 26
Affected area 27
Affected area 28
Affected area 29
Affected area 30
Affected area 31
Affected area 32
Affected area 33
Affected area 34
Affected area 35
Affected area 36
Affected area 37
Affected area 38
Affected area 39
Affected area 40
Affected area 41
Affected area 42
Affected area 43
Affected area 44
Affected area 45
Affected area 46
Affected area 47
Affected area 48
Affected area 49
Affected area 50
Affected area 51
Affected area 52
Affected area 53
Affected area 54
Affected area 55
Affected area 56
Affected area 57
Affected area 58
Affected area 59
Affected area 60
Affected area 61
Affected area 62
Affected area 63
Affected area 64
Affected area 65
Affected area 66
Affected area 67
Affected area 68
Affected area 69
Affected area 70
Company & contacts
Company & contacts
Brand: MINI, Medtronic Minimed. Contact: Medtronic MiniMed, Inc..
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.