Minimed 630g insulin pump, ref: mmt-1515, mmt-1714, mmt-1715, mmt-1754, mmt-1755; minimed 700g insulin pump, ref: mmt-1800, mmt-1801

Official notice

9 Oct 2024

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns Minimed 630g insulin pump, ref: mmt-1515, mmt-1714, mmt-1715, mmt-1754, mmt-1755; minimed 700g insulin pump, ref: mmt-1800, mmt-1801. Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA. Published 2024-10-09.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

24,595

Recall number

Z-0001-2025

Official notice

2024-10-09

The hazard

Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

95121

Recall Numbers

Z-0001-2025

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

MiniMed 630G Insulin Pump, REF: MMT-1515, MMT-1714, MMT-1715, MMT-1754, MMT-1755; MiniMed 700G Insulin Pump, REF: MMT-1800, MMT-1801, MMT-1805, MMT-1850, MMT-1851

Units affected

24,595

Hazard

Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AK

Location 2

US · United States · AL

Location 3

US · United States · AR

Location 4

US · United States · AZ

Location 5

US · United States · CA

Location 6

US · United States · CO

Location 7

US · United States · CT

Location 8

US · United States · DC

Location 9

US · United States · DE

Location 10

US · United States · FL

Location 11

US · United States · GA

Location 12

US · United States · HI

Location 13

US · United States · IA

Location 14

US · United States · ID

Location 15

US · United States · IL

Location 16

US · United States · IN

Location 17

US · United States · KS

Location 18

US · United States · KY

Location 19

US · United States · LA

Location 20

US · United States · MA

Location 21

US · United States · MD

Location 22

US · United States · ME

Location 23

US · United States · MI

Location 24

US · United States · MN

Location 25

US · United States · MO

Location 26

US · United States · MS

Location 27

US · United States · MT

Location 28

US · United States · NC

Location 29

US · United States · ND

Location 30

US · United States · NE

Location 31

US · United States · NH

Location 32

US · United States · NJ

Location 33

US · United States · NM

Location 34

US · United States · NV

Location 35

US · United States · NY

Location 36

US · United States · OH

Location 37

US · United States · OK

Location 38

US · United States · OR

Location 39

US · United States · PA

Location 40

US · United States · RI

Location 41

US · United States · SC

Location 42

US · United States · SD

Location 43

US · United States · TN

Location 44

US · United States · TX

Location 45

US · United States · UT

Location 46

US · United States · VA

Location 47

US · United States · VT

Location 48

US · United States · WA

Location 49

US · United States · WI

Location 50

US · United States · WV

Location 51

US · United States · WY

Affected area 1

US · United States

Company & contacts

Manufacturer

Medtronic MiniMed, Inc.

Phone

Not provided

Brand: MINI, Medtronic Minimed. Contact: Medtronic MiniMed, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class I

Recall details

Recall number

Z-0001-2025

Event ID

95121

Units affected

24,595

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 9 Oct 2024. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button