Open official notice

Medtronic, MyCareLink Patient Monitor, Model 24950, Rx Only. The MyCareLink Monitor, Model 24950, is a remote monitoring system that

Official notice

13 Jul 2016

RecallScope risk estimate:MediumFDA

Medtronic, MyCareLink Patient Monitor, Model 24950, Rx Only. The MyCareLink Monitor, Model 24950, is a remote monitoring system that interrogates implanted devices and transmits the data to the Care Link Network for viewing by the physician.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

262

Recall number

Z-2125-2016

Official notice

2016-07-13

The hazard

Recently, a new software version was automatically sent to a subset of Model 24950 MyCareLink Monitors. After release, Medtronic identified an issue with the software that prevents implanted device data from being available to clinicians on the CareLink Network. While the transmission appears successful to the patient, the transmitted data, including CareAlerts, are not visible to the clinic.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

74344

Recall Numbers

Z-2125-2016

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Medtronic, MyCareLink Patient Monitor, Model 24950, Rx Only. The MyCareLink Monitor, Model 24950, is a remote monitoring system that interrogates implanted devices and transmits the data to the Care Link Network for viewing by the physician.

Units affected

262

Hazard

Recently, a new software version was automatically sent to a subset of Model 24950 MyCareLink Monitors. After release, Medtronic identified an issue with the software that prevents implanted device data from being available to clinicians on the CareLink Network. While the transmission appears successful to the patient, the transmitted data, including CareAlerts, are not visible to the clinic.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AT · Austria

Affected area 4

BE · Belgium

Affected area 5

CH · Switzerland

Affected area 6

CN · China

Affected area 7

CZ · Czech Republic

Affected area 8

DE · Germany

Affected area 9

DK · Denmark

Affected area 10

ES · Spain

Affected area 11

FI · Finland

Affected area 12

FR · France

Affected area 13

GB · United Kingdom

Affected area 14

HK · Hong Kong

Affected area 15

IE · Ireland

Affected area 16

IL · Israel

Affected area 17

IN · India

Affected area 18

IT · Italy

Affected area 19

MY · Malaysia

Affected area 20

NL · Netherlands

Affected area 21

NO · Norway

Affected area 22

PL · Poland

Affected area 23

PT · Portugal

Affected area 24

SA · Saudi Arabia

Affected area 25

SE · Sweden

Affected area 26

SK · Slovakia

Affected area 27

ZA · South Africa

Company & contacts

Manufacturer

Medtronic Inc., Cardiac Rhythm and Heart Failure

Phone

Not provided

Brand: Medtronic Cardiac Rhythm And Heart Failure. Contact: Medtronic Inc., Cardiac Rhythm and Heart Failure.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2125-2016

Event ID

74344

Units affected

262

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 13 Jul 2016. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button