Open official notice

Icd-dr ddbb1d1 evera xt is1 df1 us, model number ddbb1d1; implantable cardioverter defibrillators

Official notice

28 Jun 2023

RecallScope risk estimate:CriticalFDA

ICD-DR DDBB1D1 EVERA XT IS1/DF1 US, Model Number DDBB1D1; Implantable Cardioverter Defibrillators

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

2,179

Recall number

Z-1704-2023

Official notice

2023-06-28

The hazard

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

92328

Recall Numbers

Z-1704-2023

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

ICD-DR DDBB1D1 EVERA XT IS1/DF1 US, Model Number DDBB1D1; Implantable Cardioverter Defibrillators

Units affected

2,179

Hazard

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Phone

Not provided

Brand: Medtronic Cardiac Rhythm And Heart Failure. Contact: Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF).

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class I

Recall details

Recall number

Z-1704-2023

Event ID

92328

Units affected

2,179

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 28 Jun 2023. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.