Open official notice

CareLink SmartSync Device Manager (24970A). Used to interrogate cardiac implantable electronic devices (CIED)

Official notice

12 May 2021

RecallScope risk estimate:MediumFDA

CareLink SmartSync Device Manager (24970A). Used to interrogate cardiac implantable electronic devices (CIED).

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

16,311

Recall number

Z-1522-2021

Official notice

2021-05-12

The hazard

There is a SmartSync software issue that results in an overestimation in the displayed longevity of these devices during an approximate 6-month window of time before the device triggers its Recommended Replacement Time (RRT).

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

87654

Recall Numbers

Z-1522-2021

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

CareLink SmartSync Device Manager (24970A). used to interrogate cardiac implantable electronic devices (CIED).

Units affected

16,311

Hazard

There is a SmartSync software issue that results in an overestimation in the displayed longevity of these devices during an approximate 6-month window of time before the device triggers its Recommended Replacement Time (RRT).

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Phone

Not provided

Brand: Medtronic Cardiac Rhythm And Heart Failure. Contact: Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF).

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1522-2021

Event ID

87654

Units affected

16,311

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 12 May 2021. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button