Open official notice

Hemosll ReadiPlasTin (20 ml Size), Part Number 0020301400

Official notice

14 Mar 2018

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Hemosll ReadiPlasTin (20 ml Size), Part Number 0020301400. Instrumentation Laboratory Co. Received customer reports of performance issues with some vials of Lot Nos. N1166235 and N0177760, including increased imprecision, out of range quality controls and prolonged sample results. Published 2018-03-14.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

14,894

Recall number

Z-0895-2018

Official notice

2018-03-14

The hazard

Instrumentation Laboratory Co. Received customer reports of performance issues with some vials of Lot Nos. N1166235 and N0177760, including increased imprecision, out of range quality controls and prolonged sample results.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

79165

Recall Numbers

Z-0895-2018

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Hemosll ReadiPlasTin (20 ml Size), Part Number 0020301400

Units affected

14,894

Hazard

Instrumentation Laboratory Co. Received customer reports of performance issues with some vials of Lot Nos. N1166235 and N0177760, including increased imprecision, out of range quality controls and prolonged sample results.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Instrumentation Laboratory Co.

Phone

Not provided

Brand: Instrumentation Laboratory. Contact: Instrumentation Laboratory Co..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0895-2018

Event ID

79165

Units affected

14,894

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 14 Mar 2018. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button