Acl top 300 cts 350 cts 500 cts 550 cts 970 cl: name part number 1.acl top 300 cts 00000280060 2. Acl top 300 cts 00000280060r

Official notice

4 May 2022

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Acl top 300 cts 350 cts 500 cts 550 cts 970 cl: name part number 1.acl top 300 cts 00000280060 2. Acl top 300 cts 00000280060r. Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL Q.F.A. Thrombin (Bovine), specific to instrument models that utilize a single reagent probe Published 2022-05-04.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

2,725

Recall number

Z-1014-2022

Official notice

2022-05-04

The hazard

Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL Q.F.A. Thrombin (Bovine), specific to instrument models that utilize a single reagent probe

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

89894

Recall Numbers

Z-1014-2022

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

ACL TOP 300 CTS/350 CTS/500 CTS/550 CTS/970 CL: Name Part Number 1.ACL TOP 300 CTS 00000280060 2. ACL TOP 300 CTS 00000280060R (Refurbished) 3. ACL TOP 350 CTS 00000280065 4. ACL TOP 500 CTS 00000280040 5. ACL TOP 500 CTS 00000280040R ( Refurbished) 6/ 6. ACL TOP 550 CTS 00000280045 7. ACL TOP 970 CL 00000280097

Units affected

2,725

Hazard

Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL Q.F.A. Thrombin (Bovine), specific to instrument models that utilize a single reagent probe

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

AT · Austria

Affected area 5

AU · Australia

Affected area 6

BE · Belgium

Affected area 7

BG · Bulgaria

Affected area 8

BR · Brazil

Affected area 9

CA · Canada

Affected area 10

CH · Switzerland

Affected area 11

CL · Chile

Affected area 12

CN · China

Affected area 13

CO · Colombia

Affected area 14

CR · Costa Rica

Affected area 15

CZ · Czech Republic

Affected area 16

DE · Germany

Affected area 17

ES · Spain

Affected area 18

FR · France

Affected area 19

GB · United Kingdom

Affected area 20

GR · Greece

Affected area 21

GT · Guatemala

Affected area 22

HK · Hong Kong

Affected area 23

HU · Hungary

Affected area 24

IE · Ireland

Affected area 25

IL · Israel

Affected area 26

IN · India

Affected area 27

IT · Italy

Affected area 28

JP · Japan

Affected area 29

KR · South Korea

Affected area 30

KW · Kuwait

Affected area 31

LT · Lithuania

Affected area 32

LU · Luxembourg

Affected area 33

MO · Macau

Affected area 34

MX · Mexico

Affected area 35

MY · Malaysia

Affected area 36

NL · Netherlands

Affected area 37

NO · Norway

Affected area 38

NZ · New Zealand

Affected area 39

PE · Peru

Affected area 40

PL · Poland

Affected area 41

PT · Portugal

Affected area 42

RO · Romania

Affected area 43

RS · Serbia

Affected area 44

RU · Russia

Affected area 45

SA · Saudi Arabia

Affected area 46

SG · Singapore

Affected area 47

SK · Slovakia

Affected area 48

SV · El Salvador

Affected area 49

TH · Thailand

Affected area 50

TR · Turkey

Affected area 51

TW · Taiwan

Affected area 52

UY · Uruguay

Affected area 53

VE · Venezuela

Affected area 54

VN · Vietnam

Company & contacts

Manufacturer

Instrumentation Laboratory

Phone

Not provided

Brand: Instrumentation Laboratory. Contact: Instrumentation Laboratory.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1014-2022

Event ID

89894

Units affected

2,725

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 4 May 2022. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button