Open official notice

HeartSpan Transseptal Needles Catalog No. FND-019-01

Official notice

24 Mar 2021

RecallScope risk estimate:MediumFDA

An official FDA notice concerns HeartSpan Transseptal Needles Catalog No. FND-019-01. The labeled needle tip curvature of specific lots of transseptal needles may not match the actual needle tip curvature. In the event that the incorrect needle is unknowingly used, an unintended anatomy puncture may occur, which is likely to require medical intervention. Published 2021-03-24.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

587

Recall number

Z-1262-2021

Official notice

2021-03-24

The hazard

The labeled needle tip curvature of specific lots of transseptal needles may not match the actual needle tip curvature. In the event that the incorrect needle is unknowingly used, an unintended anatomy puncture may occur, which is likely to require medical intervention.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

E1913644

Event Ids

87306

Recall Numbers

Z-1262-2021

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

HeartSpan Transseptal Needles Catalog No. FND-019-01, Lot No. E1913644 consisting of an outer needle cannula and an inner stylet. The needle is comprised of flexible thin walled tubing with an ergonomic hub and stopcock attached to the proximal end. The stylet consists of a solid wire that when inserted in the needle protrudes beyond the distal tip of the cannula.

Units affected

587

Hazard

The labeled needle tip curvature of specific lots of transseptal needles may not match the actual needle tip curvature. In the event that the incorrect needle is unknowingly used, an unintended anatomy puncture may occur, which is likely to require medical intervention.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AL

Location 2

US · United States · CA

Location 3

US · United States · CT

Location 4

US · United States · DC

Location 5

US · United States · GA

Location 6

US · United States · KY

Location 7

US · United States · LA

Location 8

US · United States · MD

Location 9

US · United States · MI

Location 10

US · United States · MT

Location 11

US · United States · NE

Location 12

US · United States · NJ

Location 13

US · United States · NY

Location 14

US · United States · PA

Location 15

US · United States · TX

Location 16

US · United States · UT

Location 17

US · United States · VA

Location 18

US · United States · WA

Affected area 1

US · United States

Company & contacts

Manufacturer

Merit Medical Systems, Inc.

Phone

Not provided

Brand: Merit Medical. Contact: Merit Medical Systems, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1262-2021

Event ID

87306

Units affected

587

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 24 Mar 2021. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button