Boston Scientific TSX Transseptal Needle, Model Number M004TSX60o; Used to create the primary puncture in the interatrial septum when

Official notice

20 Apr 2016

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Boston Scientific TSX Transseptal Needle, Model Number M004TSX60o; Used to create the primary puncture in the interatrial septum when. Merit Medical Systems, Inc. Is voluntarily conducting a recall of one lot of the TSX" Transseptal Needle due to a discrepancy between the labeling and the actual configuration of the device in the packaging. The curve dimensional values on the labels of this lot incorrectly indicate a 50¿ curve, rather than the correct 86¿ curve Published 2016-04-20.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

25

Recall number

Z-1375-2016

Official notice

2016-04-20

The hazard

Merit Medical Systems, Inc. Is voluntarily conducting a recall of one lot of the TSX" Transseptal Needle due to a discrepancy between the labeling and the actual configuration of the device in the packaging. The curve dimensional values on the labels of this lot incorrectly indicate a 50¿ curve, rather than the correct 86¿ curve

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

Q823423

Event Ids

73630

Recall Numbers

Z-1375-2016

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Boston Scientific TSX Transseptal Needle, Model Number M004TSX60o; Used to create the primary puncture in the interatrial septum when passing an intruder and/or catheter through the septum from the right side of the heart to the left side.

Units affected

25

Hazard

Merit Medical Systems, Inc. Is voluntarily conducting a recall of one lot of the TSX" Transseptal Needle due to a discrepancy between the labeling and the actual configuration of the device in the packaging. The curve dimensional values on the labels of this lot incorrectly indicate a 50¿ curve, rather than the correct 86¿ curve

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Merit Medical Systems, Inc.

Phone

Not provided

Brand: Merit Medical. Contact: Merit Medical Systems, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1375-2016

Event ID

73630

Units affected

25

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 20 Apr 2016. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button