Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Superion Indirect Decompression System
Official notice
21 Jun 2023
An official FDA notice concerns Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Superion Indirect Decompression System. Pending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver instrument tip breaks, which may result in metal fragments (Driver teeth/tips) within the implant location; and if metal fragments are not removed and remain in situ, MRI scans are NOT advised due to potential risk of patient injury. Published 2023-06-21.
What to do now
Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.
Units affected
21,533
Recall number
Z-1972-2023
Official notice
2023-06-21
The hazard
The hazard
Pending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver instrument tip breaks, which may result in metal fragments (Driver teeth/tips) within the implant location; and if metal fragments are not removed and remain in situ, MRI scans are NOT advised due to potential risk of patient injury.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Other
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Manufacturer
Phone
Brand: Boston Scientific Neuromodulation. Contact: Boston Scientific Neuromodulation Corporation.
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.