Open official notice

Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used

Official notice

27 Aug 2025

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used. Resistance and difficulties could occur while closing the retaining clip locking mechanism, which is part of a burr hole cover kit, which is part of a Deep Brain Stimulation (DBS) system, which may lead to procedural delay, or lead-migration-which may require additional surgery. Published 2025-08-27.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

24,583

Recall number

Z-2340-2025

Official notice

2025-08-27

The hazard

Resistance and difficulties could occur while closing the retaining clip locking mechanism, which is part of a burr hole cover kit, which is part of a Deep Brain Stimulation (DBS) system, which may lead to procedural delay, or lead-migration-which may require additional surgery.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

96916

Recall Numbers

Z-2340-2025

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual

Units affected

24,583

Hazard

Resistance and difficulties could occur while closing the retaining clip locking mechanism, which is part of a burr hole cover kit, which is part of a Deep Brain Stimulation (DBS) system, which may lead to procedural delay, or lead-migration-which may require additional surgery.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

AT · Austria

Affected area 5

BE · Belgium

Affected area 6

BG · Bulgaria

Affected area 7

BR · Brazil

Affected area 8

CA · Canada

Affected area 9

CH · Switzerland

Affected area 10

CL · Chile

Affected area 11

CN · China

Affected area 12

CO · Colombia

Affected area 13

DE · Germany

Affected area 14

DK · Denmark

Affected area 15

EG · Egypt

Affected area 16

ES · Spain

Affected area 17

FI · Finland

Affected area 18

FR · France

Affected area 19

GB · United Kingdom

Affected area 20

GR · Greece

Affected area 21

HK · Hong Kong

Affected area 22

HR · Croatia

Affected area 23

HU · Hungary

Affected area 24

IE · Ireland

Affected area 25

IL · Israel

Affected area 26

IN · India

Affected area 27

IS · Iceland

Affected area 28

IT · Italy

Affected area 29

JO · Jordan

Affected area 30

JP · Japan

Affected area 31

KR · South Korea

Affected area 32

KW · Kuwait

Affected area 33

LU · Luxembourg

Affected area 34

MT · Malta

Affected area 35

MX · Mexico

Affected area 36

NL · Netherlands

Affected area 37

NO · Norway

Affected area 38

PK · Pakistan

Affected area 39

PL · Poland

Affected area 40

PT · Portugal

Affected area 41

RO · Romania

Affected area 42

RS · Serbia

Affected area 43

SA · Saudi Arabia

Affected area 44

SE · Sweden

Affected area 45

SG · Singapore

Affected area 46

SK · Slovakia

Affected area 47

TW · Taiwan

Affected area 48

UA · Ukraine

Affected area 49

ZA · South Africa

Company & contacts

Manufacturer

Boston Scientific Neuromodulation Corporation

Phone

Not provided

Brand: Boston Scientific Neuromodulation. Contact: Boston Scientific Neuromodulation Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2340-2025

Event ID

96916

Units affected

24,583

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 27 Aug 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.