Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient
Official notice
20 Mar 2013
Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.
What to do now
This recall has been terminated. See the official FDA notice for the latest status.
Units affected
260
Recall number
Z-0939-2013
Official notice
2013-03-20
The hazard
The hazard
The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Location 2
Location 3
Affected area 1
Company & contacts
Company & contacts
Brand: Draeger Medical. Contact: Draeger Medical, Inc..
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.