DePuy Glenosphere orientation guide Product Usage: The glenosphere orientation guide is an instrument used in reverse shoulder

Official notice

19 Jun 2013

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns DePuy Glenosphere orientation guide Product Usage: The glenosphere orientation guide is an instrument used in reverse shoulder. The Glenosphere Orientation Guide Instrument Used with the Delta XTEND Reverse Shoulder is being recalled because an arrow that is etched on the instrument was incorrectly placed on the opposite side of the instrument. Using the improperly etched arrow as a guide could potentially lead to incorrect orientation of the glenosphere implant. Published 2013-06-19.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

20

Recall number

Z-1520-2013

Official notice

2013-06-19

The hazard

The Glenosphere Orientation Guide Instrument Used with the Delta XTEND Reverse Shoulder is being recalled because an arrow that is etched on the instrument was incorrectly placed on the opposite side of the instrument. Using the improperly etched arrow as a guide could potentially lead to incorrect orientation of the glenosphere implant.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

5120443

Event Ids

65081

Recall Numbers

Z-1520-2013

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

DePuy Glenosphere orientation guide Product Usage: The glenosphere orientation guide is an instrument used in reverse shoulder arthroplasty. The orientation guide is used to properly align the glenosphere.

Units affected

20

Hazard

The Glenosphere Orientation Guide Instrument Used with the Delta XTEND Reverse Shoulder is being recalled because an arrow that is etched on the instrument was incorrectly placed on the opposite side of the instrument. Using the improperly etched arrow as a guide could potentially lead to incorrect orientation of the glenosphere implant.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AU · Australia

Affected area 3

CA · Canada

Affected area 4

CH · Switzerland

Affected area 5

CZ · Czech Republic

Affected area 6

DE · Germany

Affected area 7

GB · United Kingdom

Affected area 8

TR · Turkey

Company & contacts

Manufacturer

DePuy Orthopaedics, Inc.

Phone

Not provided

Brand: Depuy Orthopaedics. Contact: DePuy Orthopaedics, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1520-2013

Event ID

65081

Units affected

20

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 19 Jun 2013. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button