Open official notice

Attune Revision Limb Preservation System (LPS) Tibial Insert XXSM 18MM, Part Number 151760118

Official notice

29 Mar 2023

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Attune Revision Limb Preservation System (LPS) Tibial Insert XXSM 18MM, Part Number 151760118. The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to the implant material properties. Published 2023-03-29.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

7

Recall number

Z-1263-2023

Official notice

2023-03-29

The hazard

The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to the implant material properties.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

JP6452

Event Ids

91752

Recall Numbers

Z-1263-2023

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Attune Revision Limb Preservation System (LPS) Tibial Insert XXSM 18MM, Part Number 151760118

Units affected

7

Hazard

The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to the implant material properties.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AU · Australia

Affected area 3

CA · Canada

Affected area 4

DE · Germany

Affected area 5

IN · India

Affected area 6

NZ · New Zealand

Affected area 7

PL · Poland

Company & contacts

Manufacturer

DePuy Orthopaedics, Inc.

Phone

Not provided

Brand: Depuy Orthopaedics. Contact: DePuy Orthopaedics, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1263-2023

Event ID

91752

Units affected

7

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 29 Mar 2023. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.