Open official notice

Covidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular Thin 8 mm - Long Item Code: SIGSDL45CTVT

Official notice

21 Apr 2021

RecallScope risk estimate:MediumFDA

Covidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular/Thin 8 mm - Long Item Code: SIGSDL45CTVT

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

5,928

Recall number

Z-1383-2021

Official notice

2021-04-21

The hazard

Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

N0A1038UY

Event Ids

87475

Recall Numbers

Z-1383-2021

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Covidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular/Thin 8 mm - Long Item Code: SIGSDL45CTVT

Units affected

5,928

Hazard

Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AT · Austria

Affected area 3

BE · Belgium

Affected area 4

CH · Switzerland

Affected area 5

CL · Chile

Affected area 6

CN · China

Affected area 7

DE · Germany

Affected area 8

DK · Denmark

Affected area 9

ES · Spain

Affected area 10

FI · Finland

Affected area 11

FR · France

Affected area 12

GB · United Kingdom

Affected area 13

HK · Hong Kong

Affected area 14

IE · Ireland

Affected area 15

IT · Italy

Affected area 16

NL · Netherlands

Affected area 17

NO · Norway

Affected area 18

PL · Poland

Affected area 19

PT · Portugal

Affected area 20

SE · Sweden

Company & contacts

Manufacturer

Covidien, LP

Phone

Not provided

Brand: Covidien Lp. Contact: Covidien, LP.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1383-2021

Event ID

87475

Units affected

5,928

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 21 Apr 2021. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.