Open official notice

AbsorbaTack Absorbable Fixation Device 30 Violet Absorbable Tacks 5 mm- -intended for fixation of prosthetic material to soft tissue in

Official notice

14 Feb 2024

RecallScope risk estimate:MediumFDA

AbsorbaTack Absorbable Fixation Device 30 Violet Absorbable Tacks 5 mm- -intended for fixation of prosthetic material to soft tissue in minimally invasive and open surgical hernia repair procedures Model ABSTACK30

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

54

Recall number

Z-0986-2024

Official notice

2024-02-14

The hazard

Packaging of the device may cause a low seal strength in the side seal of the foil pouch potentially resulting in a sterile/moisture barrier breach, compromising the product sterility

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

N3J1860Y

Event Ids

93722

Recall Numbers

Z-0986-2024

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

AbsorbaTack Absorbable Fixation Device 30 Violet Absorbable Tacks 5 mm- -intended for fixation of prosthetic material to soft tissue in minimally invasive and open surgical hernia repair procedures Model ABSTACK30

Units affected

54

Hazard

Packaging of the device may cause a low seal strength in the side seal of the foil pouch potentially resulting in a sterile/moisture barrier breach, compromising the product sterility

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Covidien LP

Phone

Not provided

Brand: Covidien Lp. Contact: Covidien LP.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0986-2024

Event ID

93722

Units affected

54

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 14 Feb 2024. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button