Open official notice

CardioMEMS HF System PA Sensor and Delivery System, Model Number CM2000, The CardioMEMS HF System provides pulmonary artery (PA)

Official notice

15 Mar 2023

RecallScope risk estimate:MediumFDA

An official FDA notice concerns CardioMEMS HF System PA Sensor and Delivery System, Model Number CM2000, The CardioMEMS HF System provides pulmonary artery (PA). Select CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of the intended radiofrequency has the potential to result in inaccurate readings or sensor signal acquisition difficulties. Published 2023-03-15.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

36,875

Recall number

Z-1223-2023

Official notice

2023-03-15

The hazard

Select CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of the intended radiofrequency has the potential to result in inaccurate readings or sensor signal acquisition difficulties.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

91606

Recall Numbers

Z-1223-2023

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

CardioMEMS HF System PA Sensor and Delivery System, Model Number CM2000, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which physicians use to initiate or modify heart failure treatment.

Units affected

36,875

Hazard

Select CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of the intended radiofrequency has the potential to result in inaccurate readings or sensor signal acquisition difficulties.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

AT · Austria

Affected area 5

AU · Australia

Affected area 6

BE · Belgium

Affected area 7

CA · Canada

Affected area 8

CH · Switzerland

Affected area 9

CO · Colombia

Affected area 10

CZ · Czech Republic

Affected area 11

DE · Germany

Affected area 12

DK · Denmark

Affected area 13

ES · Spain

Affected area 14

FI · Finland

Affected area 15

FR · France

Affected area 16

GB · United Kingdom

Affected area 17

IE · Ireland

Affected area 18

IT · Italy

Affected area 19

KW · Kuwait

Affected area 20

NL · Netherlands

Affected area 21

PT · Portugal

Affected area 22

SA · Saudi Arabia

Affected area 23

SE · Sweden

Company & contacts

Manufacturer

St. Jude Medical

Phone

Not provided

Brand: St Jude Medical. Contact: St. Jude Medical.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1223-2023

Event ID

91606

Units affected

36,875

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 15 Mar 2023. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button