Open official notice

BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower, REF: 352, BD Pyxis Pro

Official notice

31 Dec 2025

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower, REF: 352, BD Pyxis Pro. Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs. Published 2025-12-31.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

53,669

Recall number

Z-0962-2026

Official notice

2025-12-31

The hazard

Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

97712

Recall Numbers

Z-0962-2026

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower, REF: 352, BD Pyxis Pro 7-Drawer Auxiliary, REF: 1149-00, BD Pyxis Pro 7- Drawer Auxiliary with Secure Tower, REF: 1152-00, BD Pyxis Pro MedStation Main, REF: 1155-00, BD Pyxis CII Safe ES, REF: 1116-00

Units affected

53,669

Hazard

Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

CareFusion 303, Inc.

Phone

Not provided

Brand: Carefusion 303. Contact: CareFusion 303, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0962-2026

Event ID

97712

Units affected

53,669

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 31 Dec 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.