Open official notice

Bd pyxis medbank (name, ref): medpass module, 139088-01; mini 1fh-1fm, 169-137; mini 1hh-1fh-1hm, 169-101; mini 1hh-1hm-1fm, 169-102; mini

Official notice

30 Apr 2025

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns Bd pyxis medbank (name, ref): medpass module, 139088-01; mini 1fh-1fm, 169-137; mini 1hh-1fh-1hm, 169-101; mini 1hh-1hm-1fm, 169-102; mini. Users trying to restock a single bin location of an automated dispensing cabinet, during pocket exchange restocking workflow, may experience a software issue that results in 2 destock transactions and 2 stock transactions, which results in doubling the bin quantity, which may result in a delay in access to desired medications and/or supplies, which may lead to temporary body function impairment. Published 2025-04-30.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

7,357

Recall number

Z-1639-2025

Official notice

2025-04-30

The hazard

Users trying to restock a single bin location of an automated dispensing cabinet, during pocket exchange restocking workflow, may experience a software issue that results in 2 destock transactions and 2 stock transactions, which results in doubling the bin quantity, which may result in a delay in access to desired medications and/or supplies, which may lead to temporary body function impairment.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

96564

Recall Numbers

Z-1639-2025

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

BD PYXIS MEDBANK (Name, REF): MEDPASS MODULE, 139088-01; MINI 1FH-1FM, 169-137; MINI 1HH-1FH-1HM, 169-101; MINI 1HH-1HM-1FM, 169-102; MINI 2FH, 138902-01; MINI 2FH, 169-104; MINI 2FM, 169-105; MINI 2HH-1FH, 169-106; MINI 2HH-1FH, 138908-01; MINI 2HH-1FM, 169-107; MINI 2HH-1FM, 138906-01; MINI 3HH-1FG, 169-110; MINI 3HH-1FG, 139091-01; MINI 3HH-1HM, 169-111; MINI 4HH, 138910-01; MINI 4HH, 169-112; MINI 4HM, 169-113; MINI 4HM, 138915-01; TWR MN 10HH-1FH-1FM, 169-23; TWR MN 10HH-2FH, 169-24; TWR MN 10HH-2FH, 138955-01; TWR MN 10HH-2FM, 169-25; TWR MN 10HH-2FM, 138924-01; TWR MN 12HH-1FF, 169-26; TWR MN 12HH-1FF, 138960-01; TWR MN 12HH-1FH, 169-27; TWR MN 12HH-1FH, 138961-01; TWR MN 12HH-1FM, 169-28; TWR MN 12HH-1FM, 138926-01; TWR MN 14HH, 138927-01; TWR MN 14HH, 169-29; TWR MN 1HH-1HM-2FM-4FF, 169-30; TWR MN 2FH-2HM-4FM, 169-31; TWR MN 2HH-10HM-1FM, 169-32; TWR MN 2HH-1FH-5FM, 169-33; TWR MN 2HH-2FH-4FM, 169-34; TWR MN 2HH-4FH-2FM, 169-139; TWR MN 2HH-4HM-4FM, 139125-01; TWR MN 2HH-6FM, 169-38; TWR MN 2HH-6FM, 138919-01; TWR MN 2HH-6HM-3FM, 169-39; TWR MN 2HH-8HM-2FM, 169-40; TWR MN 2HM-6FM, 169-41; TWR MN 3FH-4FM, 169-42; TWR MN 3FH-4FM, 138933-01; TWR MN 3FH-8HH MEDPASS, 169-140; TWR MN 3HH-7HM-2FM, 169-43; TWR MN 4HH-10HM, 169-44; TWR MN 4HH-10HM, 138959-01; TWR MN 4HH-1FH-2HM-3FM, 169-45; TWR MN 4HH-1FH-4FM, 169-46; TWR MN 4HH-1FH-4HM-2FM, 169-47; TWR MN 4HH-2FH-2HM-2FM, 169-49; TWR MN 4HH-2FH-3FM, 169-50; TWR MN 4HH-2FH-3FM, 138931-01; TWR MN 4HH-3FH-2FM, 169-52; TWR MN 4HH-4FH-1FM, 169-53; TWR MN 4HH-4FH-1FM, 138940-01; TWR MN 4HH-4HM-3FM, 169-54; TWR MN 4HH-4HM-3FM, 138952-01; TWR MN 4HH-5FH, 169-55; TWR MN 4HH-5FM, 169-56; TWR MN 4HH-5FM, 138918-01; TWR MN 4HH-6HM-2FM, 169-57; TWR MN 4HM-5FM, 169-59; TWR MN 5HH-3HM-3FM, 169-60; TWR MN 5HH-9HM, 138914-01; TWR MN 5HH-9HM, 169-61; TWR MN 6FH-1FM, 169-62; TWR MN 6HH-1FH-2HM-2FM, 169-63; TWR MN 6HH-1FH-3FM, 169-64; TWR MN 6HH-2FH-2FM, 169-65; TWR MN 6HH-4FH, 169-67; TWR MN 6HH-4FH, 138957-01; TWR MN 6HH-4FM, 169-68; TWR MN 6HH-4FM, 138922-01; TWR MN 6HH-4HM-2FM, 169-69; TWR MN 6HH-8HM, 169-71; TWR MN 6HH-8HM, 138958-01; TWR MN 7FH, 169-72; TWR MN 7FH, 138920-01; TWR MN 7FM, 169-73; TWR MN 7FM, 138937-01; TWR MN 7HH-1HM-3FM, 139086-01; TWR MN 7HH-1HM-3FM, 169-74; TWR MN 8HH-1FH-2FM, 169-75; TWR MN 8HH-2FH-1FM, 169-77; TWR MN 8HH-2HM-2FM, 169-78; TWR MN 8HH-2HM-2FM, 138939-01; TWR MN 8HH-3FF, 169-79; TWR MN 8HH-3FH, 169-80; TWR MN 8HH-3FH, 138956-01; TWR MN 8HH-3FM, 169-81; TWR MN 8HH-3FM, 138923-01. BD PYXIS MEDFLEX MN (Name, REF): 1000 12HH, 139055-01; 1000 2HH-2FM, 139041-01; 1000 2HH-5FM, 139046-01; 1000 4HH-4FM, 139049-01; 1000 6HH-3FM, 139051-01; 1000 8HH-2FM, 139052-01; 2000 8HH-2FH, 139165-01; 2000 8HH-2FM, 139069-01. MEDBANK MAIN, 1137-00; MEDBANK MINI MAIN, 1147-00; MEDBANK TOWER MAIN, 1145-00; MEDFLEX, 1119-00; MEDFLEX 2.0, 1139-00

Units affected

7,357

Hazard

Users trying to restock a single bin location of an automated dispensing cabinet, during pocket exchange restocking workflow, may experience a software issue that results in 2 destock transactions and 2 stock transactions, which results in doubling the bin quantity, which may result in a delay in access to desired medications and/or supplies, which may lead to temporary body function impairment.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

CareFusion 303, Inc.

Phone

Not provided

Brand: Carefusion 303. Contact: CareFusion 303, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1639-2025

Event ID

96564

Units affected

7,357

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 30 Apr 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button