Open official notice

Alinity i Processing Module, 03R65 01, with the below components. A. TED (Thermo Electric Device) Engine, Reagent cooler; b. CABLE, POWER

Official notice

5 Jun 2019

RecallScope risk estimate:HighFDA

An official FDA notice concerns Alinity i Processing Module, 03R65 01, with the below components. A. TED (Thermo Electric Device) Engine, Reagent cooler; b. CABLE, POWER. Potential loose cable connections on the reagent cooler, which could result in temperature errors and eventual reagent cooler failure. Could cause burn and shock injuries to users and a delay in test results. Device design has mitigated this risk, but it is a possibility. Published 2019-06-05.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

373

Recall number

Z-1497-2019

Official notice

2019-06-05

The hazard

Potential loose cable connections on the reagent cooler, which could result in temperature errors and eventual reagent cooler failure. Could cause burn and shock injuries to users and a delay in test results. Device design has mitigated this risk, but it is a possibility.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

82595

Recall Numbers

Z-1497-2019

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Alinity i Processing Module, 03R65 01, with the below components. a. TED (Thermo Electric Device) Engine, Reagent cooler; b. CABLE, POWER OUT, TED CNTRLER Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples.

Units affected

373

Hazard

Potential loose cable connections on the reagent cooler, which could result in temperature errors and eventual reagent cooler failure. Could cause burn and shock injuries to users and a delay in test results. Device design has mitigated this risk, but it is a possibility.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

AT · Austria

Affected area 5

AU · Australia

Affected area 6

BE · Belgium

Affected area 7

BR · Brazil

Affected area 8

CA · Canada

Affected area 9

CH · Switzerland

Affected area 10

CL · Chile

Affected area 11

CN · China

Affected area 12

CO · Colombia

Affected area 13

CZ · Czech Republic

Affected area 14

DE · Germany

Affected area 15

ES · Spain

Affected area 16

FI · Finland

Affected area 17

FR · France

Affected area 18

GB · United Kingdom

Affected area 19

GR · Greece

Affected area 20

HK · Hong Kong

Affected area 21

HR · Croatia

Affected area 22

IL · Israel

Affected area 23

IN · India

Affected area 24

IT · Italy

Affected area 25

JP · Japan

Affected area 26

KR · South Korea

Affected area 27

MX · Mexico

Affected area 28

NL · Netherlands

Affected area 29

NO · Norway

Affected area 30

PK · Pakistan

Affected area 31

PL · Poland

Affected area 32

PT · Portugal

Affected area 33

RO · Romania

Affected area 34

RS · Serbia

Affected area 35

SA · Saudi Arabia

Affected area 36

SE · Sweden

Affected area 37

TH · Thailand

Affected area 38

TR · Turkey

Affected area 39

VN · Vietnam

Affected area 40

ZA · South Africa

Company & contacts

Manufacturer

Abbott Gmbh & Co. KG

Phone

Not provided

Brand: Abbott Kg. Contact: Abbott Gmbh & Co. KG.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1497-2019

Event ID

82595

Units affected

373

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 5 Jun 2019. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button