Open official notice

Alinity c Carbon Dioxide Reagent Kit, Produced for Abbott by Sekisui Diagnostics P.E.I. Inc. Prince Edward Island Canada

Official notice

4 Dec 2019

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Alinity c Carbon Dioxide Reagent Kit, Produced for Abbott by Sekisui Diagnostics P.E.I. Inc. Prince Edward Island Canada. Abbott internal studies have determined that atmospheric Carbon Dioxide (CO2) can be absorbed into the Alinity c Carbon Dioxide reagent, resulting in the potential for incorrect results. Abbott was made aware of this issue from customer complaints. Internal testing demonstrated that the amount of CO2 absorbed is higher with increased reagent carousel rotation and when the volume of reagent in the cartridge is reduced. This phenomenon can be detected as a shift in QC. The shift is predictive in both direction and magnitude of the potential impact to patient results. Published 2019-12-04.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

105

Recall number

Z-0575-2020

Official notice

2019-12-04

The hazard

Abbott internal studies have determined that atmospheric Carbon Dioxide (CO2) can be absorbed into the Alinity c Carbon Dioxide reagent, resulting in the potential for incorrect results. Abbott was made aware of this issue from customer complaints. Internal testing demonstrated that the amount of CO2 absorbed is higher with increased reagent carousel rotation and when the volume of reagent in the cartridge is reduced. This phenomenon can be detected as a shift in QC. The shift is predictive in both direction and magnitude of the potential impact to patient results.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

54454UQ03, 54449UQ03

Event Ids

84140

Recall Numbers

Z-0575-2020

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Alinity c Carbon Dioxide Reagent Kit, Produced for Abbott by Sekisui Diagnostics P.E.I. Inc. Prince Edward Island Canada

Units affected

105

Hazard

Abbott internal studies have determined that atmospheric Carbon Dioxide (CO2) can be absorbed into the Alinity c Carbon Dioxide reagent, resulting in the potential for incorrect results. Abbott was made aware of this issue from customer complaints. Internal testing demonstrated that the amount of CO2 absorbed is higher with increased reagent carousel rotation and when the volume of reagent in the cartridge is reduced. This phenomenon can be detected as a shift in QC. The shift is predictive in both direction and magnitude of the potential impact to patient results.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AU · Australia

Affected area 4

BE · Belgium

Affected area 5

BR · Brazil

Affected area 6

CA · Canada

Affected area 7

CH · Switzerland

Affected area 8

CO · Colombia

Affected area 9

DK · Denmark

Affected area 10

ES · Spain

Affected area 11

FR · France

Affected area 12

GB · United Kingdom

Affected area 13

GR · Greece

Affected area 14

HK · Hong Kong

Affected area 15

IN · India

Affected area 16

IT · Italy

Affected area 17

NL · Netherlands

Affected area 18

NZ · New Zealand

Affected area 19

PK · Pakistan

Affected area 20

PL · Poland

Affected area 21

PT · Portugal

Affected area 22

RO · Romania

Affected area 23

SA · Saudi Arabia

Affected area 24

SE · Sweden

Affected area 25

SG · Singapore

Affected area 26

TH · Thailand

Affected area 27

ZA · South Africa

Company & contacts

Manufacturer

Abbott GmBH & Co. KG

Phone

Not provided

Brand: Abbott Kg. Contact: Abbott GmBH & Co. KG.

Source & status

Source

Tier

official

Category

ELECTRONICS

Official classification

Class II

Recall details

Recall number

Z-0575-2020

Event id

84140

Units affected

105

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 4 Dec 2019. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button