Open official notice

AIA-360 Automated Immunoassay Analyzer, Product Code 019945

Official notice

11 Mar 2020

RecallScope risk estimate:MediumFDA

AIA-360 Automated Immunoassay Analyzer, Product Code 019945

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

12

Recall number

Z-1420-2020

Official notice

2020-03-11

The hazard

A display screen software issue on the analyzer causes the display to freeze when display screen is touched at the same time as a command from the instrument firmware, causing the instrument to stop. As a result, the run is aborted and the results are not retrievable.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

84787

Recall Numbers

Z-1420-2020

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

AIA-360 Automated Immunoassay Analyzer, Product Code 019945

Units affected

12

Hazard

A display screen software issue on the analyzer causes the display to freeze when display screen is touched at the same time as a command from the instrument firmware, causing the instrument to stop. As a result, the run is aborted and the results are not retrievable.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AZ

Location 2

US · United States · NE

Location 3

US · United States · NJ

Location 4

US · United States · NY

Location 5

US · United States · WA

Affected area 1

US · United States

Affected area 2

CR · Costa Rica

Affected area 3

UY · Uruguay

Company & contacts

Manufacturer

Tosoh Bioscience Inc

Phone

Not provided

Brand: Tosoh Bioscience. Contact: Tosoh Bioscience Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1420-2020

Event ID

84787

Units affected

12

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 11 Mar 2020. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button