Open official notice

Access Testosterone assay, Catalog Number 33560

Official notice

11 Jun 2025

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Access Testosterone assay, Catalog Number 33560. Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results. Published 2025-06-11.

What to do now

Do not consume or use the affected product. Return it to the place of purchase or follow instructions in the official FDA notice.

Units affected

70

Recall number

Z-1901-2025

Official notice

2025-06-11

The hazard

Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

538081

Other

UDI-DI 15099590231347

Event Ids

96824

Recall Numbers

Z-1901-2025

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Access Testosterone assay, Catalog Number 33560

Units affected

70

Hazard

Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · FL

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

VN · Vietnam

Company & contacts

Manufacturer

Beckman Coulter, Inc.

Phone

Not provided

Brand: Beckman Coulter. Contact: Beckman Coulter, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1901-2025

Event ID

96824

Units affected

70

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 11 Jun 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button